Skip to end of metadata
Go to start of metadata

You are viewing an old version of this page. View the current version.

Compare with Current View Page History

« Previous Version 412 Next »


Refer to the Material and Process Compatibility page for information on materials compatible with this tool.
Equipment Status
: Set as UP, PROBLEM, or DOWN, and report the issue date (MM/DD) and a brief description. Italicized fields will be filled in by BNC Staff in response to issues. See Problem Reporting Guide for more info.

Status

DOWN

Issue Date and Description

 Extended electrical shutdown necessary for JEOL 2 install

Estimated Fix Date and Comment

 System should be available and calibrated by 5 PM.

Responding Staff

Wirth, Justin C 

iLab Name

JEOL JBX-8100FS E-Beam Writer

iLab Kiosk

BRK Lithography Core

FIC

Shared

Owner

Bill Rowe and Justin Wirth (Email, Chat)

Location

BRK 2100P

Max. Wafer

6"/150 mm

Info Links

SOP/wiki/spaces/BNCWiki/pages/6232822 | /wiki/spaces/BNCWiki/pages/6232164

Overview

General Description

 Click here to expand...

The new JEOL JBX-8100FS series spot beam lithography system is designed for higher throughput and lower operating costs. The JBX-8100FS writes ultrafine patterns at a faster rate of speed while minimizing idle time, especially during the exposure process, thus increasing throughput. This new, high precision compact e-beam tool is suitable for a wide range of applications from research to production, while its small footprint and low power consumption reduce cost of ownership.

Main Features

  • Small footprint
    The area required for the standard system is 4.9 m (W) x 3.7 m (D) x 2.6 m (H), much smaller than the conventional systems.

  • Low power consumption
    Power needed for normal operation is approximately 3 kVA, reduced to 1/3 of the conventional systems.

  • High throughput
    The system has two exposure modes, high resolution and high throughput modes, supporting different types of patterning from ultra fine processing to small to mid size production. It has minimized the idle time during exposure while increasing the maximum scanning speed by 1.25 to 2.5 times to 125 MHz (the world’s highest level) for high speed writing.

  • Version
    The JBX-8100FS is available in 2 versions: G1 (entry model) and G2 (full option model). Optional accessories can be added to the G1 model as needed.

  • New Functions
    An optional optical microscope is available to enable examination of patterns on the sample without exposing resist to light. A signal tower is provided as standard for visual monitoring of system operation.

  • Laser positioning resolution
    Stage positions are measured and controlled in 0.6 nm steps as standard, and in 0.15 nm steps with an optional upgrade.

  • System control
    Versatile Linux® operating system combined with a new graphic user interface provides ease in operation. The data preparation program supports both Linux® and Windows®.

Specifications

 Features

The JBX-8100FS is an electron-beam lithography system designed to write nanometer to submicron sized patterns using a spot beam.

Write Modes Specifications

Note: High Throughput will be almost exclusively used, contact Justin Wirth if you think you need HR mode.


High Throughput
(HT/4th Lens/EOS 3)

High Resolution
(HR/5th Lens/EOS 6)

Max. Main Field Size

1000 μm x 1000 μm

100 μm x 100 μm

Max. Sub Field Size

8 μm x 8 μm

0.8 μm x 0.8 μm

Min. Beam Step Size

0.5 nm

0.05 nm

Min. Shape Placement Step

1 nm

0.1 nm

Overlay Accuracy

≤±20 nm

≤±9 nm

Field Stitching Accuracy

≤±20 nm

≤±9 nm

Min. Beam Diameter

5.1 nm

1.8 nm

Min. Line Width (Field Center)

<12 nm

<8 nm

Available Currents (at BNC)

2, 10, 30, 100 nA (60 nA available as a test file)

0.5 nA

Features

  • ZrO/W emitter 

  • 4-stage electron-beam focusing system

  • Accelerating voltage: 100 kV

  • Writing: Vector scan (within a subfield) and Step-and-repeat (electronically between subfields and physically between fields).

  • Beam scanning speed: ≤125 MHz

  • Scan speed modulation: 256 rank / 0.05 nsec resolution

  • Mainfield/Positioning DAC: 20-bit

  • Subfield/Scanning DAC: 14-bit

  • Focus range: ±100 μm

  • Max wafer size: 200 mm

  • Max writing area: 150 mm x 150 mm

  • Movable area: 190 mm x 170 mm

  • Stage positioning resolution: λ/1024 (~0.6 nm)

  • Beam current stability: 0.2% pp/hr

  • Beam position stability: ≤60 nm pp/hr (HT) / ≤10 nm pp/hr (HR)

  • Substrate thickness compatibility: 225 μm to 1.3 mm

  • Deflection amplitude correction and objective-lens focus correction, using the substrate height detector

  • Smallest features size: 4.2 nm (demonstrated by JEOL in Development of the JBX-8100FS Electron Beam Lithography System).

Error rendering macro 'excerpt-include' : User 'null' does not have permission to view the page 'JEOL JBX-8100FS - Internal Resources - Detailed Specifications'.

The JBX-8100FS is mounted on a TMC Quiet Island with STACIS III antivibration supports. Transmission is minimized at high frequencies, and unlike older anti-vibration supports, is reduced at low (<10 Hz) frequencies as well.

Sample Requirements and Preparation

Samples need to be free of outgassing contaminants, and PR must be properly baked to avoid contamination of the column

Standard Operating Procedure

SOP - JEOL JBX-8100FS E-Beam Writer

Process Control Information

Process Control Context

 Process Control Information Context

The JEOL system is calibrated by staff every 7 - 21 days, with separate calibration necessary for each condition file (current). These are typically stable for 1-3 weeks. As part of this process, column shift and tilt is adjusted (similar to an SEM), current is measured (and adjusted via the column zoom lenses, if necessary), wobble is checked and minimized (via the objective aperture, again similar to an SEM), focus and astigmatism is adjusted (via the objective lens strength and stigmator correction values), and the DAILYCAL file is run to ensure everything passes.

Data is shown for the current FEG and previous FEG. These are replaced approximately every 2 years.

Some particularly relevant measurements are shown below:

  • Current: specified as nominal values. As part of the exposure, the machine will measure the current (by default, every 20 minutes) and automatically adjust the exposure dose based on this. As such, drifts in current will not directly affect the writing quality, but may affect user experience on the system.

    • How is this set?

      • Current is a result of the particular characteristics of the emitter at a given point in its lifetime (which may increase or decrease current at the sample), the staff set values of shift and tilt (which should maximize it), and most importantly, the values of the zoom lenses in the column. Current is adjusted by adjusting the zoom lenses. This is not done frequently due to the needed time for accurate adjustment, and the minimal impact on write quality that current drift has for users. Current settings are adjusted whenever a particular condition file is getting close to the specification limit.

    • How is this measured?

      • The system moves to the Faraday cup and takes, takes a number of instantaneous measurements of the current, and averages them together.

    • Upper specification limit: the nominal current itself

      • Via the dose equation, too high of a staff set current may cause the machine to need a clock speed higher than 125 MHz if a user chooses a typical clock speed very close to the limit (e.g. typically >=120 MHz). This would result in an error message when trying to create the magazine file. From the user standpoint, this could only be remedied by increasing the minimum dose (not usually desirable, as this may overdose your pattern) or increasing the shot pitch (which may or may not result in different pattern results, but is also generally not desirable because of the unknown effect on your results). As a result, staff endeavor to keep the actual current below the nominal current at all times.

    • Lower specification limit: 10% less than the nominal current

      • Too low of a current (within reason) will not in anyway harm the pattern, but the write will take longer than expected. If the write is limited by the beam on time, 10% lower current (90% of the specified current) will take ~11% longer to write (100%/90% ≅ 1.11). This is considered a reasonable error, but longer than ~11% may unexpectedly prolong the write, and thus staff will endeavor to keep the actual current not lower than 10% lower than the nominal current.

  • Calibration measured beam size: The beam size in the X and Y directions measured by the system on a “clean” spot of the AE mark used by staff for calibration. This will always be larger than the true size of the beam, and is only useful in the context explained here.

    • How is this set?

      • This is set by having proper alignment by staff of the blanking aperture, shift, tilt, focus, and astigmatism. Additionally, as a particular mark is used to for measurement of the beam size, it will get “contaminated”, artificially inflating the measured beam size. Staff will slightly move the measured position every week, but there is still some low level inflation of the measured beam size that may not be real due to this mark contamination.

    • How is this measured?

      • The beam is scanned over a metal knife edge on top of a current detector, which does math to obtain the size of the beam, assuming it is Gaussian. There are scans in both the X and Y direction. The Gaussian assumption is a good assumption at low beam currents (< 30 nA) and an increasingly poorer assumption at high beam currents (> 30 nA), so this value is only given for low currents. Note that the values here will be lower than values given when the focus program is run as part of DAILYCAL because a less clean mark is used for that (which in no way means the beam is actually larger, nor does it negatively affect the write quality). A less pristine mark position is used for writes vs. calibrations because the machine will still properly focus even if the mark is slightly contaminated. The mark positions are moved as needed by staff to ensure a clean enough mark is used at all times.

    • Upper specification limit: theoretical beam size plus an inherent scattering factor due to the measurement technique plus an allowable blur.

    • Lower specification limit: none.

  • Astigmatism:

    • How is this set?

      • The focus program is run in a different mode, which measures the beam size at different focus and stigmator values, and find the minimum size across these different values.

    • How is this measured?

      • The astigmatism value returned is the number of DAC points between the X and Y best focus values per a curve fit.

    • Upper specification limit: + 20 DAC points.

    • Lower specification limit: - 20 DAC points.

High currents (>30 nA) have their focus and astigmatism set manually by staff rather than through the built-in calibration routines on the system. Similarly, the built-in measurement programs do not give reliable values for beam size or astigmatism. As a result, only beam current (which is measured similarly and reliably compared to low currents) is posted here.

Process Control Charts

 2 nA

 10 nA

 30 nA

 High Current - 60 nA and 100 nA

Process Library

Currently empty, please contribute your processes here.

8100calculator.xlsx

Use this Excel file to assist with picking currents, shot pitches, ensuring you're within the clock headroom (<125 MHz, >8ns), to roughly estimate your write time based on the current/dose/pattern area, and track alignment mark locations.

8100calculator.xlsx

 Kiosk Links


iLab Kiosk

eLog: Submit

eLog: View/Edit

Report Problem


 Login to view link

Error rendering macro 'excerpt-include' : User 'null' does not have permission to view the page 'eLog - JEOL JBX-8100FS E-Beam Writer'.



 Login to view link


References

BNC Staff Presentations - Internal Resources

J. C. Wirth, "Introduction to the JEOL JBX-8100FS", BNC Faculty Seminar, 2/08/2018 (Updated 10/25/2018).

Purdue's JBX-8100FS, from JEOL - Internal Resources

JEOL Manuals - Internal Resources

 Click here to expand...

General References

Relevant Literature

 Click here to expand...

Partnership Opportunities - Alternate EBL Resists/Processes

 Click here to expand...

Contact Justin Wirth if you are interested in partnering with BNC to evaluate these resists and develop standard processes of broad usefulness to the BNC research community. 

AZ nLOF 2000

AllResist

AR-P 6200 and AR-N 7520 are of particular interest:

FAQ: E-Beam Resists

AR-P 6200:

Nanostructures

Developer Comparison

High Contrast Developer

  • No labels