iLab Name: Olympus Microscope
iLab Kiosk: BRK Characterization Core
FIC: Shared
Owner: Bill Rowe
Location: BRK 2100
Maximum Wafer Size:
Overview
General Description
The Olympus BX-60 Microscope can be used for sample inspection in reflection mode. Image capture is obtained though an Insight (18.2 Color Mosaic) CCD camera and "Spot Advanced" software.
Specifications
- Reflection Mode Light Source: 100Watt halogen
- Available observations:
- Reflected light brightfield
- Reflected light darkfield
- Reflected light Nomarski DIC
- Reflected light simplified polarized light
- Objectives: 5X, 10X, 20X, 50X
- Stage Travel: 105 x 100 x 52 mm
Tool Overview
Nomenclature / Identification | |||||||||||||||||||||
Controls | |||||||||||||||||||||
Eye Piece
| |||||||||||||||||||||
BF/DF Light Path Selection
Effects of Dummy Slider -The dummy slider (2) has been installed on the revolving nosepiece at the factory. It can be replaced by loosening the clamping knob (3) when a DIC prism is to be used. However, in observations other than DIC, be sure to push in the dummy slider in order for prevention of flare. | |||||||||||||||||||||
Filters ND Filter Knob - The ND filter is interlocked with the brightfield (BF) light path switching so it can be engaged or disengaged according to the mirror selector lever (1). The ND filter makes it possible to reduce the glare when darkfield (DF) is switched to brightfield (BF). Filter Slots - Engage the optimum filter sliders for the purpose of observation in the two filter insertion slots (3). Be sure to engage from the left side. The first click position is the idle position and the second click engages the filter in the light path.
| |||||||||||||||||||||
Intensity Ratio Selector Push or pull the intensity ratio selector to select the desired light path.
| |||||||||||||||||||||
Field Iris The field iris is used to restrict the diameter of the beam of light entering the objective and thus excludes stray light from the observation path improving image contrast. The diameter of the field iris should be adjusted for each objective power to the extent that is just circumscribes the field of view. The field iris diaphragm knob (2) and the field iris centering screw (3) are used to adjust the field iris size and position in relation to the observation path. | |||||||||||||||||||||
Aperture Iris The aperture iris diaphragm determines the numerical aperture of the illumination system. Matching the numerical aperture of the illumination system with that of the objective provides better image resolution and contrast, and also increases the depth of focus. The aperture iris diaphragm knob (2) and the aperture iris centering knob (3) are used to adjust the aperture iris size and position in relation to the observation path. | |||||||||||||||||||||
Revolving Nosepiece The revolving nosepiece allows the user to easily swap objective lenses into the observation path. Grasp the knurled ring and rotate the nosepiece clockwise or counterclockwise until you feel the detent snap into position. If the objective is not located at a detent with objective installed, then the observation path will be blocked and the user will only see black. Do not attempt to rotate the nosepiece by grasping the object lense assembly directly. This can damage the objective lense. | |||||||||||||||||||||
Illumination Controls
| |||||||||||||||||||||
Focus Knob
| |||||||||||||||||||||
XY Axis Knob
|
Standard Operating Procedure
General Observation Procedure for Reflected Mode Microscopy | |||||
---|---|---|---|---|---|
Set the cube selector (1) to BF for bright field. | |||||
Disengage the filter slider, analyzer slider, and polarizer slider (3) from the light path. | |||||
Make sure light intensity preset button (3) is OFF. | |||||
Press the main switch (3) to "I" (ON) and adjust the light intensity with the light intensity knob (1). | |||||
Using the coarse focus adjustment knob (1), move the stage to the lowest position. Place sample on the stage. Note: Be sure that the surface of the specimen to be observed is perpendicular to the optical axis of the microscope. | |||||
Turn the revolving nosepiece to engage the 5X objective. While looking through the eyepiece, use the coarse focus adjustment knob to raise the stage and bring the sample into focus. | |||||
Adjust interpupillary distance as necessary for clear binocular vision. | |||||
Looking only through the right eyepiece with your right eye, use the fine adjustment knob to bring the sample into final focus. | |||||
Looking only through the left eyepiece, use the diopter ring (1) to bring the left eyepiece into focus. This corrects the microscope observation paths for any differences between your left and right eye. You should now see a clear image through both lenses. | |||||
Adjust the field iris as necessary to obtain appropriate illumination for the sample.
Reflected light bright field observation - To obtain good image contrast, adjust the diameter of the illuminating beam in accordance with the objective in use. Using the field iris diaphragm knob (2) on the vertical illuminator, adjust the diaphragm so that the field of view is circumscribed by the field iris diaphragm in order to exclude stray light. Reflected light dark field observation - Always keep the field iris diaphragm knob (2) pushed in to leave the diaphragm open. | |||||
Adjust the aperture iris as necessary to obtain appropriate illumination for the sample.
Reflected light bright field observation - In general, a good image is obtained if the diaphragm is reduced down to 70-80% of the objective's numerical aperture. Reflected light dark field observation - Generally, you will always keep the aperture iris diaphragm knob (2) pushed in to leave the diaphragm open. Note: Depending on the specimen, an image with good contrast and little flare may sometimes be obtained by keeping the aperture iris diaphragm stopped down a little more. Please experiment with this to see if it works with the particular specimen. | |||||
Engage the filters suitable for your observation
| |||||
Photomicrography | |||||
This microscope is equipped with a photomicrography tube and camera for observation and photography using the attached PC. |
Frequently Asked Questions
- Question 1?
- Answer 1.