Overview

General Description

Ozone generator supplying ozone to the Fiji ALD and Black Magic Systems, located in GH Chase of the cleanroom, consists of two main components.  

  • a ODS-9000 Series high-purity ozone generator - The ozone generator contains ozone generating equipment and controls including an oxygen and nitrogen MFCs, an in-line ozone concentration sensor (gFFOZ).
    • Image Added
  • a SCI-TURBO-L controller - The SCI-TURBO-L controller connects to the ozone generator and controls the flowrate of each process gas, ensures safety interlocks are met, and displays ozone concentrations.
    • Image Added
Tip

This is an example of a tip


Warning

Use of controls, adjustments, or procedures other than those specified in this manual without consulting a competent safety professional may result in exposure to potential hazards.


Warning

Ozone (O3) is a toxic gas. High concentrations of ozone are dangerous and harmful to humans. Take reasonable steps to avoid exposure. The current maximum 8-hour exposure limit for ozone is 0.1 PPM (according to U.S. OSHA). Concentrations above 0.1 ppm can cause headaches, eye irritation, respiratory problems, dizziness, and nausea. At much higher concentrations it can cause death.


Warning

Non-flammable materials may become flammable in the presence of high oxygen and/or ozone concentrations!


Warning

The ozone monitor contains an ultraviolet (UV) lamp. Never remove the lamp from its housing while the unit is powered on. Eye exposure to UV light is extremely dangerous.


Warning

High voltages are used in the process of generating ozone, however all circuits of the generator are either covered or isolated.

Specifications

Oxygen Flow:  0.25 to 5 slpm

Ozone Outlet Pressure: Internally set to 30 PSIG

Ozone Outlet Flow Range: 0-5000 sccm

Standard Operating Procedure

Starting the Ozone delivery system

Unless the ozone delivery system is currently being used, you will find the system is not operating and gases are not flowing.  Identified by the following initial state of the tool.

Sci-Turbo L Controller Display

Image Added

MFC RD Screen shows 0 on all values


Generator

Image Added

Display shows 0 % power

Knob should be turned all the way CCW until it stops.

Ozone Switch is in the off position

Toggle switch is at Local and indicated by the green light

Red light indicating Ext. Interlock conditions not met


Image Added

Press the D button to advance the menu to "MFC Parameters"

Image Added

Press the E button to access "MFC N Parameters"

Image Added

Press the circle Circle button to advance to access "MFC A Parameters"

Image Added

Press E to access the menuSet to 5 VDCthe setting


Image Added

Press the left arrow 3 times to activate the first whole number (0.00 VDC)

Press the up arrow until 5 is displayed (5.00 VDC)

Press the E button to input 5.00 VDC as the set point

Image Added

Red Ext. Interlock light should shut off and green standby light will come on.

Verify flow by advancing to MFC RD Screen

Flip Ozone Switch to On position

Rotate Knob CW to increase the power

Set dosage rate