Etching a Sample |
---|
Activating the System- Use the kiosk to make your logbook entry.
- Log into iLab and start your reservation and activate the tool.
| |
Check System Status- If the system goes in to a gauge calibration routine, the total time is just over one hour.
- It is best to let the system complete the gauge calibration. Contact the staff and they can adjust your time to remove the gauge calibration charges.
- Do not use if currently in use or maintenance signage is present.
|
|
Logging onto the System- When starting the Xactix system the prompt to the right will be displayed.
- Enter the username and password below.
- Username: user
- Password: user
| |
Loading / Unloading a Sample- Press the Load/Unload Sample sample button on the Main menu screen.
- The system will launch a verification window. Press YES yes to continue, NO to abort chamber vent.
- Image Added
- Verification prompt is provided since the purge/vent process can be time consuming and inconvenient if accidentally started.
- The system window changes and begins chamber Load/Unload Sample process window will load and begin 3 chamber pump / purge cycles to evacuate the chamber.
- When the chamber is vented, the dialog box below will appear and you can open the chamber lid. The lid will rest open on the stop behind chamber.
- Load / Unload your sample and close the lid.
- Press Examine or Done button.
- Examine button - Pumps the chamber down quickly, without purges, so that the system can be quickly vented again to load the sample. This is useful when examining a sample away from the system, and prevents moisture from accumulating in the chamber. In general better results are obtained by etching in one run rather than several shorter runs where the sample is removed and then replaced.
- Done button - The system will go through a purging cycle prior to chamber pump-down. It is always necessary to press Done before etching the sample.
Note |
---|
Do not leave the chamber lid open longer than necessary. Any moisture accumulated will negatively affect the etch process. |
- After pressing Done, and the chamber pumps down, it will go back to the main screen, and the Machine Status information box at bottom left will show “Ready”.
Note: Verification prompt is provided since the purge/vent process can be time consuming and inconvenient if accidentally started- .
| |
- Select Etch Menu button from the Main Menu Screen.
- A prompt asking for a lot number will appear.
- Enter an alpha-numerical lot number which will be recorded with your etch data. This entry does not affect your process in any way.
- Press the Done button
- The Etch Menu will show on screen. The name of last used recipe is shown on top left.
| |
Load an Existing Recipe- Click the current recipe pull down to expand.
- Select the desired recipe.
- Verify the # of Cycles, Etch Time, XeF2 Pressure, and N2 pressure values are correct for your process.
| |
Editing Process Parameters- Set the four process parameters for your etch process.
- # of cycles - The depth of etching is controlled by the number of cycles. A cycle consists of the xenon difluoride sublimating to the set pressure in the expansion chamber, opening to chamber and etching for a set amount of time and evacuation of the main chamber and expansion chambers.
- Etch Time - The etch time This is essentially the exposure time between the opening of of the sample to XeF2 etchant. More precisely the etch time counter starts when the valve between the charged expansion chamber and the evacuated process chamber and the opening of opens, and stops when the valve between the process chamber and the vacuum pump . This is essentially the exposure time of the sample to XeF2 etchant.Note: opens.
- Although it is sample dependent, it has been found that XeF2 XeF2 is typically fully depleted after 30 seconds of exposure to the Si sample.
- You can estimate the depletion time by watching the chamber pressure after exposure to XeF2 etchant. Pressure will rise to a stable value once XeF2 has been consumed.
- Xef2 Pressure -
- N2 Pressure -
Image RemovedSince the XeF2 reaction with Si produces more moles of gas (2XeF2 + Si -> SiF4+ 2Xe) as it reacts, you will see the chamber pressure rise as the reaction continues (keep in mind that you may see a normal leakup of the chamber also). When the available XeF2 has been consumed, the pressure in the chamber should not stop rising. If you watch the Chamber Pressure carefully throughout an exposure, you should see a point where the etch chamber pressure essentially stops rising. You can use this time, plus 10%, to set your cycle time since any additional time is essentially wasted.
- Xef2 Pressure - To introduce a consistent amount of XeF2 into the main chamber, a set pressure charge of XeF2 is delivered to the expansion chamber before exposing the sample to the etchant. XeF2 has a vapor pressure of ~4T at 25 C.
- Assuming all XeF2 is consumed in each cycle, this parameter determines etch depth per cycle.
- Typical values used for this parameter at our facility are between 1 - 2.5 Torr.
- It is recommended to set the XeF2 pressure below 3 Torr. The system must stabilize the expansion chamber pressure at this set point before proceeding. Setting higher pressures may result in very slow cycle times, or failure to proceed.
- N2 Pressure - XeF2 is added to the expansion chamber first, followed by N2. In this way, an accurate and reproducible mixture of gas can be delivered to the sample.
- The addition of N2 improves the selectivity to silicon nitride.
- The addition of large amount of nitrogen can dilute the etchant and help achieve very slow etch rates.
- Etch Mode - Select between Normal and Normal with delay under the Etch Mode pull-down.
- Normal with delay allows the user to add a delay time between cycles so the wafer can cool down.
| Image Added |
Saving a Recipe- To Save a new recipe base on current settingsbased on currently displayed process parameters, click the Save button. The save new recipe window will appear.
Global Recipes are starting recipe templates available to all users, please do not overwrite these. Use the toggle switch in the middle of the screen to set to Personal Recipes. Users should save their new recipe to this location. Type in the name of your new recipe and Click on the SAVE button. You will be returned to the Etch Menu Screen. Please start the file name with your Purdue alias. (example: jshepar-Recipe1)
| Image Added Image Added |
Run a Recipe- From the Etch Menu Screen, and after etch recipe and parameters have been verified, press the Start Etch button.
Progress of the etch can be monitored using the tools in the lower right corner of the Etch Menu screen. - When the etch completes normally, the system will load the Main Menu screen and you will see the message Etch completed, press for details displayed and flashing.
| Image Added Image Added |
Stopping an Etch Before Recipe Completes- From the in progress Etch Menus Screen, the Stop button is used to end an etch recipe prematurely.
- Pressing the Stop button once initiates a soft stop. The tool will complete the current cycle and then stop executing the recipe and return to the Main Menu screen.
- Pressing the Stop button twice initiates a dialog box asking for input.
- Image Added
- Pressing Yes, Hard Stop will stop the recipe immediately.
- Pressing No, Keep Waiting will revert to a soft stop described above.
| Image Added |
Changing the Number of Cycles During an EtchUnloading a Sample- Pressing the CHANGE CYCLES button at any time during an in progress etch will prompt the user with the Change # of Cycles menu.
- The dialog shows # of Cycles remaining currently.
- The top row of arrows will add cycles in this order, (right to left); ones, tens, hundreds, and thousands.
- The lower row of arrows, having the same values, will remove cycles.
- Shows the result if changes are applied.
- Press OK to apply changes, Cancel to exit without changing.
Example: In the image to the right, An extra 125 cycles have been added to the etch by clicking the right most arrow (ones) five times, the next right most arrow (tens) twice, and the 2nd from left most arrow (hundreds) once. | Image AddedImage Added |
| Finishing |
Creating or Changing a Recipe |
Etch Recipe Parameters |