Overview

General Description

The Olympus BX-60 Microscope can be used for sample inspection in reflection mode. Image capture is obtained though an Insight (18.2 Color Mosaic) CCD camera and "Spot Advanced" software.

Specifications

  • Reflection Mode Light Source: 100Watt halogen
  • Cube Assemblies: Bright-Field (BF), Dark-Field (DF), and Differential Interference Contrast (DIC).
  • Objectives: 5X, 20X, 50X

Tool Overview

Nomenclature / Identification



Controls

Image Added

*The above illustration shows the microscope with the U-RLBC mirror cube housing for the brightfield/darkfield vertical illuminator.  In the case of Birck's microscope, U-URBC is used and the cube selector knob is replaced by a turret.

*Not all controls are available on the Birck microscope.

Image Added

Illumination Controls

  1. Light Intensity Adjustment Slider
  2. Light Intensity / Voltage Indicator
    1. For photomicrography the intensity should at approximately the level indicated by the camera symbol (Image Added)
    2. The numbers to the right indicate the approximate voltage of the light.
  3. Transmitted / Reflected light selector switch.
    1. Image Added

Image Added

Light Path Selection

  1. Select the cube according to the observation purpose.
    1. Bright Field
    2. Dark Field


Image Added

Filters

Pushing or pulling filters will place them into or out of the observation path.

  2. Filter Slider

  3.  Analyzer Slider

  4. Polarizer Slider


Image Added

Image Added

Revolving Nosepiece

The revolving nosepiece allows the user to easily swap objective lenses into the observation path.  Grasp the knurled ring and rotate the nosepiece clockwise or counterclockwise until you feel the detent snap into position.

If the objective is not located at a detent with objective installed, then the observation path will be blocked and the user will only see black.

Warning

Do not attempt to rotate the nosepiece by grasping the object lense assembly directly.  This can damage the objective lense.


Image Added

Eye Piece

  1. Interpupillary distance adjustment
    1. While looking through the eyepieces, adjust for binocular vision until the left and right fields of view coincide completely.  You should see one clear image when properly adjusted, out of adjustment and you will see double images.
  2. Diopter Adjustment
    1. Looking through the right eyepiece (2) with your right eye, rotate the course and fine adjustment knobs to bring the specimen into focus.
    2. Looking through the left eyepiece with your left eye, turn the diopter adjustment ring (1) to focus on the specimen.
  3. Eye Shades
    1. Since you will be using safety glasses of some sort, use with the eye shades in their normal folded down position. This will prevent the eyeglasses from being scratched.

Image Added

Image Added

Image Added

Focus Knob
XY Axis Knob
Field Iris
Aperture Iris
General Observation Procedure for Reflected Mode Microscopy



Sample Requirements and Preparation


Standard Operating Procedure


Questions & Troubleshooting