Location: Cleanroom GH Chase

Overview

General Description

Ozone generator supplying ozone to the Fiji ALD and Black Magic Systems, located in GH Chase of the cleanroom.

This is an example of a tip

This is an example of a tool warning

Specifications

Range: 0-5000 sccm

Standard Operating Procedure

Starting the Ozone delivery system

Unless the ozone delivery system is currently being used, you will find the system is not operating and gases are not flowing.  Identified by the following initial state of the tool.

Sci-Turbo L

MFC RD Screen shows 0 on all values

Generator

Display shows 0 % power

Knob should be turned all the way CCW until it stops.

Ozone Switch is in the off position

Toggle switch is at Local and indicated by the green light

Red light indicating Ext. Interlock conditions not met


Press the D button to advance the menu to "MFC N Parameters"

Press the circle button to advance to "MFC A Parameters"

Press E to access the menu

Set to 5 VDC


Red Ext. Interlock light should shut off

Verify flow by advancing to MFC RD Screen

Flip Ozone Switch to On position

Rotate Knob CW to increase the power

Set dosage rate