Using the Glovebox: General Guidelines
1. The most obvious thing: NEVER put anything containing oxygen (closed vial would be a good example) or water/moisture into the glovebox.
2. ALWAYS record all significant events—catalyst regeneration, nitrogen tank exchange, any malfunctions, changes of pump oil, changes of filters, replacements of glovebox parts—in the provided logbook, along with your name and the date of the event.
3. NEVER leave both the inner and the outer door to the antechamber open.
4. When in doubt—pump it out! If you are not sure whether your antechamber is under nitrogen or vacuum, ALWAYS pump it out. If you are not using it, leave the antechamber under
vacuum.
5. When working with substances with very fine particle sizes, and you are afraid they could
spill all over the place, turn off the CIRCULATOR SWITCH during your manipulation.
Make sure to turn it back on once you are done.
6. There is very little space available in the glovebox, and it must be used effectively. Common
chemicals are organized into three large boxes, with (a) main group metals, (b) transition metals, and (c) organic compounds. Each user should organize their own smaller container to keep their samples in it. NEVER borrow things without asking.
7. Before putting your hands into the gloves, ALWAYS remove your watch, rings, and bracelets. This will minimize the risk of making holes in the gloves.
8. ALWAYS put on new clean nitrile or latex gloves before you enter the glovebox gloves. The
concern is purely hygienic. The atmosphere within the glovebox is about 3–7 °C warmer than the outside (because of the circulator’s hot motor), so you will sweat. Use some baby powder (provided on top of the glovebox) to minimize this discomfort.
9. ALWAYS be mindful of the pressure. This is especially important if you are operating the antechamber or putting your hands in or out of the box. The pressure should not be too low (gloves sucked into the box), or too high (gloves sticking straight up from the box). Use pedatrol to adjust the pressure differences quickly; automatic pressure control will eventually kick in. If it does not, something is wrong: contact Karolina or Ognjen immediately.
10. All vessels brought into the glovebox must be nitrogen-filled or evacuated before. If you do not do this, you will introduce air into the box. Make sure that all empty containers are open. When putting in a vessel with a solvent—especially a volatile one, like Et2O—or a solution, the stopper must be properly secured, even on flasks which were evacuated before. If not, it will very likely pop out and make a big mess in the antechamber. Use plastic clips and additional rubber bands to hold the stopper in place.
11. All solvents or solutions that you plan to put into the box must be dry and degassed.
12. The glovebox catalyst can be irreversibly damaged by thiols, amines, phosphines, alkyl halides, SO2, SO3, Hg, etc. You should avoid these materials in the glovebox at all cost. If you absolutely have to use them, ALWAYS discuss your experiment in detail with Ognjen
before trying something potentially harmful.
13. In the glovebox, keep your samples open only when necessary. Do not leave any open vials
in the box after you are done, since they will be thrown away.
14. When weighing materials in the glovebox, static electricity is a big problem. To minimize
this disturbance, handle all materials that you plan to weigh using the provided tongs—not your gloves directly! Use the provided antistatic gun to "shoot" at the balance, tongs, and all of your vials/vessels before you place them on the balance.
15. When you are done using the box for the day, ALWAYS clean the box up! Leave it in good condition, remove all your waste. The accumulated mess in the glovebox is really hard to clean.
Bringing Materials into the Glovebox
1. Make sure that the antechamber inner door is closed.
2. If the antechamber is evacuated, close the valve to the pump.
3. Open the nitrogen valve leading to the box slowly to fill the antechamber with nitrogen. Once
this is done (observe the pressure gauge), close the valve again.
4. Open the outer door, place your materials onto the antechamber shelf, and close the outer
door again.
5. SLOWLY open the antechamber vacuum pump valve.
6. Place the antechamber under vacuum and keep it under vacuum for at least 20 minutes. If you are bringing in a lot of material, extend this time to 30 minutes. If you are bringing in very porous materials—paper, cork rings, kimwipes, etc.—you first evacuation cycle must last overnight.
7. Close the antechamber pump valve and slowly fill it with nitrogen again.
8. Repeat steps 5–7 two more times.
9. Put your hands into the gloves, open carefully the inner door, and bring in your materials.
10. Closetheantechamberinnerdoor.
Bringing Materials out of the Glovebox
1. Make sure that the antechamber is under nitrogen. If you are not sure about this, you must evacuate the antechamber and then refill it with nitrogen. Do not make any assumptions!
2. Open the inner door, put your stuff in, and close the inner door tightly.
3. Make sure that the valve leading from the antechamber to the glovebox is closed. Make sure
that the antechamber vacuum valve is closed too.
4. Open the outer door, remove your stuff, and close the door again.
5. Slowly open the antechamber vacuum valve. Leave the antechamber under vacuum.
Catalyst Regeneration and Atmosphere Testing
The inert atmosphere in glovebox is purified by circulating it through a catalyst bed (also known as DRI-TRAIN). The catalyst bed contains a copper-based oxygen-reducing agent and molecular sieves to remove water. The catalyst bed needs to be regenerated periodically, depending on the usage and on the types of materials that were used in the glovebox. This period is typically on the order of 3–6 months. Catalyst regeneration is performed by: (a) isolating the catalyst from the glovebox, (b) exposing the catalyst to a stream of hydrogen and heating it, (c) evacuation of the catalyst, and (d) refilling of the catalyst with inert gas and reconnecting it to the box. The mixture of 5% H2 in N2 (uncertified) is used as a regeneration gas. One full tank is roughly enough for two regeneration cycles. During the regeneration, the chemically bound oxygen is released as water and expelled from the catalyst bed—together with the moisture released from the molecular sieves—through the stream of the regeneration gas. The full regeneration cycle takes about 12 hours: it should be started early in the morning and closely followed until the evening. Do not perform the regeneration process overnight!