Overview
General Description
The Olympus BX-60 Microscope can be used for sample inspection in reflection mode. Image capture is obtained though an Insight (18.2 Color Mosaic) CCD camera and "Spot Advanced" software.
Specifications
- Reflection Mode Light Source: 100Watt halogen
- Available observations:
- Reflected light brightfield
- Reflected light darkfield
- Reflected light Nomarski DIC
- Reflected light simplified polarized light
- Objectives: 5X, 10X, 20X, 50X
- Stage Travel: 105 x 100 x 52 mm
Tool Overview
Nomenclature / Identification | |||||||||||||
Controls | |||||||||||||
Eye Piece
| |||||||||||||
BF/DF Light Path Selection
| |||||||||||||
Filters There are two filter sliders (1) at the end of the illuminator. Two filters can be set on each filter slider. The desired filters can be brought into the optical path by sliding the filter sliders in and out.
| |||||||||||||
Intensity Ratio Selector Push or pull the intensity ratio selector (1) to select the desired light path.
| |||||||||||||
Field Iris The field iris is used to restrict the diameter of the beam of light entering the objective and thus excludes stray light from the observation path improving image contrast. The diameter of the field iris should be adjusted for each objective power to the extent that is just circumscribes the field of view. The field iris diaphragm slider (1) and the field iris centering screw (2) are used to adjust the field iris size and position in relation to the observation path. | |||||||||||||
Aperture Iris The aperture iris diaphragm determines the numerical aperture of the illumination system. Matching the numerical aperture of the illumination system with that of the objective provides better image resolution and contrast, and also increases the depth of focus. The aperture iris diaphragm slider (1) is used to adjust the aperture iris size in relation to the observation path. Typically adjusted to 7-80% of the objective's exit pupil. | |||||||||||||
Revolving Nosepiece The revolving nosepiece allows the user to easily swap objective lenses into the observation path. Grasp the knurled ring and rotate the nosepiece clockwise or counterclockwise until you feel the detent snap into position. If the objective is not located at a detent with objective installed, then the observation path will be blocked and the user will only see black.
| |||||||||||||
Illumination Controls
| |||||||||||||
Focus Knob
| |||||||||||||
Coarse Focus Stopper The coarse focus stopper restricts the movement of the coarse focus knob so the stage cannot be raised beyond the operators specified limit. With the sample in focus, turn the coarse focus stopper ring as far as it will go in the direction of the arrow on the base of the microscope (about 3/4 revolution). The coarse focus stopper is now clamped in position. When changing the sample, lower the stage by turning only the coarse focus knob. After changing the sample, gently raise the stage by turning only the coarse focus knob as far as it will go. The sample should be roughly in focus when the stage has been raised as far as it will go. Use the fine focus knob to bring the sample into perfect focus. |
Standard Operating Procedure
General Observation Procedure for Bright-field Reflected Mode Microscopy |
---|
Set the cube selector (1) to BF for bright field.
Press the main switch ( |
1) to "I" (ON) |
Using the coarse focus adjustment knob (1), move the stage to the lowest position. Place sample on the stage.
Note: Be sure that the surface of the specimen to be observed is perpendicular to the optical axis of the microscope.
Adjust the field iris as necessary to obtain appropriate illumination for the sample.
- Pull out the field iris diaphragm knob (2) to reduce the diameter of the diaphragm.
- If necessary, turn the centering screw (3) to adjust so that the image of the diaphragm is centered in the field of view.
- To check center, push the field iris knob (2) in to increase the diameter of the diaphragm until it touches the periphery of the field of view. If the image is not centered, adjust as necessary.
- Further enlarge the field iris diaphragm diameter until its image just circumscribes the field of view.
Reflected light bright field observation - To obtain good image contrast, adjust the diameter of the illuminating beam in accordance with the objective in use. Using the field iris diaphragm knob (2) on the vertical illuminator, adjust the diaphragm so that the field of view is circumscribed by the field iris diaphragm in order to exclude stray light.
Reflected light dark field observation - Always keep the field iris diaphragm knob (2) pushed in to leave the diaphragm open.
Adjust the aperture iris as necessary to obtain appropriate illumination for the sample.
- Remove the eyepieces to observe the diaphragm directly.
- Looking through the eyepiece sleeves, pull out the aperture iris diaphragm knob (2) to leave the diaphragm approximately 70% open.
- If necessary, center the diaphragm by manipulating the aperture iris diaphragm centering screw (3).
Reflected light bright field observation - In general, a good image is obtained if the diaphragm is reduced down to 70-80% of the objective's numerical aperture.
Reflected light dark field observation - Generally, you will always keep the aperture iris diaphragm knob (2) pushed in to leave the diaphragm open.
Note: Depending on the specimen, an image with good contrast and little flare may sometimes be obtained by keeping the aperture iris diaphragm stopped down a little more. Please experiment with this to see if it works with the particular specimen.
Engage the filters suitable for your observation
Usable Filters | Applications |
---|---|
U-BP1100IR (Band Pass Filter) | Infrared |
. | |
Configure the microscope for Bright-field microscopy
| |
Place sample on the microscope stage, and bring it into focus while looking through the binocular eyepieces. | |
Adjust the diopter. | |
Adjust the interpupillary distance. | |
Change the magnification and observe the sample.
Tip: When observing a sample with small contrast, such as the mirror-faced sample, it can be rather difficult to focus on it. In such a case, stop down the field diaphragm so that its image can be seen in the eyepieces. Now attempt to focus on the rim of the diaphragm image. When the rim is in focus, the sample will also be in focus. | |
Dark-field Microscopy | |
Focus on the sample using bright-field microscopy techniques discussed above. | |
Configure the microscope for dark-field microscopy.
Note: The field and aperture diaphragms are fully-opened automatically. However, the lever positions do not change. | |
Photomicrography | |
This microscope is equipped with a photomicrography tube and camera for observation and photography using the attached PC. |
Frequently Asked Questions
- Question 1?
- Answer 1.