LEFT: schematic of the inductively coupled plasma (ICP) argon ion source, user sample (red) is placed to the right of the accelerator grid. “Neut.” Refers to the neutralizer. RIGHT: schematic of the Hiden Ion Milling Probe using Secondary Ion Mass Spectrometry (SIMS) for etch End Point Detection (EPD). 

Page Properties
idInfo


iLab NameC - AJA ICP Argon Ion Mill Etcher
iLab KioskBRK Etch Core
FICSunil Bhave
OwnerNick Glassmaker
LocationCleanroom - Q Bay
Max. Wafer6"/150 mm
Info Links/wiki/spaces/BNCWiki/pages/6251296/wiki/spaces/BNCWiki/pages/6238484Internal | Staff



Tool Highlights

  • Direct physical milling of thin and thick films using argon atoms.
  • Ion beam source : inductively-coupled argon plasma, accelerated up to 900 V in a 14cm broad beam with currents up to 620 mA (Kaufman & Robinson, Inc.).
  • Ar+ ions neutralized using a matched electron beam current before reaching the sample.
  • water-cooled rotating sample carrier can be oriented 0-90 degrees from normal beam incidence.
  • Secondary ion mass spectrometer (SIMS) mounted on the system can be used for manual or automatic endpoint detection (EPD) of milled material by discerning elements based on charge/mass ratio of positive ions (Hiden Analytical).
  • Sample load-lock system saves time with processing.
Page Properties
idEtch_Capabilities


TypeMaterialsRestricted MaterialsAvailable GasesMax RF PowerWafer Size
ICP Ion Milloxides, metals, polymersAuAr900 W6"




iLab KioskeLog: SubmiteLog: View/EditReport Problem


Expand
titleLogin to view link

Insert excerpt
eLog - AJA ICP Argon Ion Mill Etcher
eLog - AJA ICP Argon Ion Mill Etcher
nopaneltrue



Expand
titleLogin to view link

Image RemovedImage Added




Page Properties
idFunctionality


StatusUP


Estimated Fix Date and Comment


Responding StaffNeilNick G.




Table of Content Zone
locationtop
styledisc

Table of Contents
outlinetrue
indent25px
stylenone


Sample Requirements and Preparation

The shared carrier and mask wafers are dirty since they collect all the sputtered material from users' samples.

If you care about contamination of your sample due to re-sputtering of material, you need to provide your own carrier and masks.

About etching gold (Au): please discuss with Neil or the BNC staff in charge before etching Au. 

Sample chuck cooling during etch: Note that thick films require high etch rates (and hence high amounts of heating), so this may mean one needs to load the sample directly on the chuck by venting the chamber (instead of using load lock). This will permit users to apply more torque and get better cooling of the chuck. However, it requires special training.

For EPD, note that argon atomic mass is the same as MgO, so the EPD is not able to sense (ionized) MgO in the large Ar ion background.

Standard Operating Procedure

View file
nameSOP - AJA ion mill.pdf
height250


Process Library

See Internal Resources page. 


Questions & Troubleshooting



References

Presentation: Introduction to AJA Ion Mill (Fab Forum, Feb 2019)

View file
nameKRI AC DISCHARGE CONTROLLER MANUAL AC1006_V2.pdf
height250
View file
nameKRI Auto Controller Manual RFAC1 LFN Manual.pdf
height250
View file
nameKRI Emission Controller Manual EC12002_V2.pdf
height250
View file
nameKRI Ion Optics Controller Manual 9007-0007_IOC15XX_01_VD.PDF
height250
View file
nameKRI LFN2000 NEUTRALIZER MANUAL_S_VB.PDF
height250
View file
nameKRI RF grid 7117-Certificate.pdf
height250
View file
nameKRI RFICP 140 ION SOURCE and ion optics MANUAL 9007-RF-1401_VC.pdf
height250
View file
nameMKS 937B gauge controller manual.pdf
height250
View file
nameSeren RF power supply manual_RLX01_V2_02.pdf
height250
View file
nameEBARA TMP MANUAL ME_PU_EMT2200MK/3300MK-B_STD(CN)_R0 sec pa.pdf
height250
View file
nameIMP End Point Detector User Manual.pdf
height250
View file
nameMASsoft Professional Manual IMP-EPD.pdf
height250
View file
nameAJA logbook.pdf
height250
View file
nameVAT gate valve series 11.pdf
height250
View file
nameMKS 974 gauge Manual - REV C.PDF
height250