Overview

Sample cleanliness is extremely important for achieving good results in nanofabrication (hence the need for a cleanroom). However, simply working in the cleanroom does not guarantee cleanliness, proper steps need to be taken at every processing step to ensure a clean result. While some processes have more stringent cleanliness needs than others, some general rules apply for all types of lithography:

  • Handle substrates/wafers with cleaned tweezers, not hands.
  • Substrates should not be left exposed. Although the cleanroom offers nearly particle free air, particles are generated by the movement, plastic wafer carrier, talking, etc. Even in the cleanest areas of the cleanroom, particles will eventually find their way onto exposed surfaces.
 


Solvent Cleaning

Standard Process: TAI clean

  1. 5 min sonication in beaker of Toluene.
  2. 5 min sonication in beaker of Acetone.
  3. 5 min sonication in beaker of Isopropanol.

Background

Standard solvent cleaning is a three step process of ultrasonically agitating the substrate in subsequent solvent solutions to remove organic contamination. Three steps allow for selecting solvents that properly dissolve multiple types of contaminants. The first step is a nonpolar solvent (toluene) to remove any oils or grease present on the sample. The second step is a powerful polar solvent (acetone) to remove polar contaminants such as photoresist. The final step is a polar protic solvent (isopropanol).

Nonpolar Solvent Clean: Toluene

Toluene acts a nonpolar solvent to dissolve any grease or oil on the substrate. Nonpolar contaminants (especially fingerprints) will not be effectively removed by Acetone or IPA.  (Interesting Fact:  Toluene is used in the process to remove cocaine from coca leaves in the production of Coca-Cola syrup!!)

Polar Aprotic Solvent Clean: Acetone

Acetone's high dipole moment leads to a powerful cleaning of polar residue (such as photoresist). However its quick evaporation leads to residue redepositing on the substrate if evaporation is allowed to occur.

Polar Protic Solvent Clean: Isopropanol

IPA is used to remove fully wash away acetone from the substrate, and remove any remaining dissolved residue. As a result, IPA is mainly used to removing the Acetone without residue. However, unlike Acetone, IPA will form hydrogen bonds, and thus may provide some additional cleaning effects.

IPA is prefered as a final step over methanol due to its .  Its significantly slower evaporation rate . Slower evaporation results in less residue , and the and a more forgiving rinse step. The close dipole moments of the two chemicals (DIPA=1.66 vs. DMethanol=1.69) means the difference between the two in cleaning ability is small. Additionally, although methanol is somewhat stronger. From a safety standpoint, unlike Methanol, IPA is not considered toxic. And importantly for cleaning, "IPA is generally the purest available organic solvent and is used extensively for vapor drying of H2O-rinsed wafers." (Reinhardt 2008, p. 22)

If the clean is done immediately before spinning photoresist, a thorough IPA clean and dry should dehydrate the wafer, removing the need for a dehydration bake. If there is a significant delay between cleaning and spinning, ambient water will be adsorbed on the wafer surface. A dehydration bake should then be done.

Deionized Water Rinse

Note

A final DI rinse is usually optional, but is required for samples to be placed in furnaces.

A powerful final step may be a rinse in the nanopure deionized water, which will assist in the removal of any remaining particles, solvent residues, and acid/base residues. However, if this step is employed before spinning photoresist, a dehydration bake must be performed to properly remove bound surface water molecules. This step is required for samples to be placed in any high temperature equipment or furnaces, as even slight solvent residue will combust at temperatures above 400 C.

 


Acid Cleaning Processes

Standard Process: Piranha Clean

  1. A completely dry glass container should be used.
  2. Add 3 parts Sulfuric Acid (H2SO4).
  3. Add 1 part Hydrogen Peroxide (H2O2).
  4. Allow the solution to work for 5-10 minutes.
  5. With sink running and while flushing with large amount of water, slowly pour used Piranha solution down drain.

Background

Piranha is a strong oxidizing agent that will strip surfaces of organic (carbon containing) matter and hydroxylate surfaces (add OH groups, making the surface hydrophilic).

  • Samples should be free of large areas of organics. Piranha should not be used for removing a layer of PR off of a sample, a solvent clean should be used for large amounts of organics. Piranha is appropriate for cleaning off any remaining organic residue after a solvent clean, or for cleaning a new wafer.
  • Ratios of H2SO4 and H2O2 are important. Getting close to the 3:1 ratio (3 parts H2SO4: 1 part H2O2) is important. Ratios down to 7:1 will still give powerful cleaning, but something within the range of 3:1 to 7:1 should be used. Adding more peroxide (e.g. a 2:1 mixture or 1:1) both presents a safety hazard and weakens the strength of the solution, and should not be done.
  • Contrary to the conventional rule of "Always Add Acid" in chemistry, Piranha should always be prepared with acid first, and peroxide slowly added second. Improperly prepared Piranha may bubble violently when adding the chemicals, this is avoided by adding chemicals slowly and adding the sulfuric acid first.
  • Water should NOT be added to Piranha solution and ALL beakers, tweezers and samples added to Piranha must be completely dry. Piranha is self heating, and can easily reach temperatures above 100 C. The addition of water can cause violent boiling.
  • Piranha solution should be left uncovered, and under no circumstances should it be sealed. The oxygen and oxidation byproducts produced by the reaction are flammable, and sealing the container can cause them to explode.

Piranha alternative: Nanostrip

Birck supplies Nanostrip, a stabilized Piranha alternative. Standard Nanostrip consists of 90% sulfuric acid (H2SO4), 5% peroxymonosulfuric acid (H2SO5), 5% water (H2O), and <1% hydrogen peroxide (H2O2). A more powerful version, Nanostrip 2X, consists of 85% sulfuric acid (H2SO4), 10% peroxymonosulfuric acid (H2SO5), 5% water (H2O), and <1% hydrogen peroxide (H2O2). However, to be effective

 

Questions & Troubleshooting

 

Process Library

References

Solvents: Theory and Application (via


Characterizing Cleanliness

Particle inspection (Particle size ~0.5 um and larger)

Inspection for contamination by particles 0.5 um or larger can be achieved readily through dark field microscopy.  

In bright field illumination, the object is lit from below the stage, resulting in a larger, contrasted image that can be studied.

A dark field microscope is ideal for viewing objects that are unstained, transparent and absorb little or no light.  These specimens often have similar refractive indices as their surroundings, making them hard to distinguish with other illumination techniques.  A dark field microscope blocks this central light with a condenser so that only oblique rays hit the object.  If there is nothing on the stage, the aperture of the condenser is greater than the objective and the view will be completely black.  It is more useful in examining external details, such as outlines, edges, grain boundaries and surface defects than internal structure.

Image result for dark field microscope wafer inspectionImage Added

https://www.microscopeworld.com/t-darkfield_microscopy.aspx


Image resultImage Added

Dark-field microscope (a) and top-view scanning electron microscope (SEM) (b) images of the same area on a silicon wafer ablated by a femtosecond laser. Microscope image is inverted in horizontal direction relative to that of the SEM. Selected nanoparticles are marked by corresponding numbers 1 to 6 in both figures.

https://www.researchgate.net/figure/Dark-field-microscope-a-and-top-view-scanning-electron-microscope-SEM-b-images-of_fig4_228331975


Residual Metal Characterization

Metals present on a wafer can be quantitively characterized though a number of techniques, though we do not (to our knowledge) readily have the capabilities at Purdue to perform these measurements. Outside companies can be enlisted to perform ICP-MS / VPD ICP-MS to characterize what metals may be present on a sample. This is more cost effective than technologies such as TXRF, which show where on the wafer the contamination is rather than simply how much metal is present across the entire wafer.