Page Properties
idFunctionality

Insert excerpt
Problem Reporting Guide
Problem Reporting Guide
nopaneltrue

StatusUP
Issue Date and Description


Estimated Fix Date and Comment

Responding Staff

Image Removed




iLab Name: C - ASAP-1 IPS Digital Sample Preparation System
iLab Kiosk: BRK Packaging and Assembly Core
FIC:
Walter Leon-Salas

Page Properties
idFunctionality
Insert excerptProblem Reporting GuideProblem Reporting GuidenopaneltrueStatusDOWNIssue Date and Description

 Spindle speed is lagged in grind step

Estimated Fix Date and Comment
Responding StaffJoon Park

Image Removed

iLab Name: C - ASAP-1 IPS Digital Sample Preparation System
iLab Kiosk: BRK Packaging and Assembly Core
FIC:
Walter Leon-Salas
Owner: Jerry Shepard

Owner: Jerry Shepard

Location:
 BRK 2221
Maximum Wafer Size: 
4"/100 mm

Table of Content Zone
locationtop
styledisc

Table of Contents
outlinetrue
indent25px
stylenone

Overview

Expand

Note: the following is quoted from ULTRA TEC's documentation

General Description

ASAP-1® IPS is a digital sample preparation system for the decapsulation, thinning and polish­ing of packaged wafer-level devices.

Drawing on ULTRA TEC’s market knowledge and long-standing leadership in electronic sample preparation, IPS is ‘device centric’ – designed to meet the non-destructive, high yield and survivability needs of micro-electronic components.

IPS’ Touch Screen OS has been designed from the ground-up to control all aspects of the sample preparation process. Advanced programming drives IPS’s deep sub-micron axes of motion. The machine vision monitor provides an unique ‘always-live’ image of the device, along with key navigational and process information.

The touch-off tool control provides Z-position at pre-defined locations and pressures – enabling package and die tilt to be corrected quickly and accurately. Tool height variations are automatically-calibrated.

Selected area preparation

"Selected Area Preparation (SAP) has long been established as a low damage machining method, with the use of relatively low spindle speeds, in conjunction with the floating head providing very low temperature rises. This is in stark contrast with high speed 'mills' which introduce excess vibration, ultrasonics, and process temperature rises -- generally requiring either a cooling bath to be in position at all time, or even a constant high flow of coolant to avoid device overheating. SAP is almost always a ‘wet process’ . However the coolant required is minimal."

The use of ULTRA TEC’s patented 'float down' head approach has enabled several hundred customers to achieve high quality polished surfaces, ready for analysis. High yields have been shown using analog ASAP-1® equipment for thinning silicon to less than 20 micron target RST. The move to higher accuracy digital systems allows for routine thinning of extremely thin substrates at even low single-digit RST’s. The move to larger flip chips mounted on PCB substrates requires inherent design for 3D sample prep. Sample Preparation developments have been driven by the industry's push for both smaller RST’s – for optimal backside imaging and attenuation – and the need for improved uniformity, to satisfy the latest analytical techniques.

• “A Straightforward Guide to the Sample Preparation of Curved & Warped IC's”; TEC Note #16, 2015 • “Backside Failure Analysis Techniques: What’s The Gain Of Silicon Getting Thinner”; Boit, Schafer et al; ESREF 2014 • “The State of the Art in Backside Sample Prep”; Jim Colvin, EDFA Vol 4, No.2, 2002 • “Packages Have Become the New IC’s”; Jim Colvin, EDFA Volume 16, Issue 4, November 2014

Specifications

  • Suits all sizes of die - package, wafer and board-level
  • Real Time Video Monitor with system parameter
  • Touchscreen control with physical joystick & controls
  • Rigidized Table Assembly and closed-loop, high-torque, motor control enhances the machining of tough and hard materials
  • X, Y and Z axes all have deep sub-micron accuracy
  • Accurately decaps, then thins substrate and polishes
  • Patented Floating Head provides a true polishing action -- yields polishing quality & high survivability
  • Intuitive menus provide a powerful, easy to use, system
  • USB Flash Driveinterface for preparation recipe storage
  • Short set-up and process times
  • Accurate die-tilt adjustment ‘on the fly’
  • Bench-top & Quiet in Operation
Z-Vertical Direction Precision0.04 microns (40 nanometers)Table Precision (X& Y Travel)0.2 microns (200 nanometers)Table Travel Amplitude100mm x 100mmPolishing MethodPatented ASAP-1 Float-down head, with Z-lock, enhanced with electronic sensors and tool patternsVideoReal-time machine vision with 6.5 inch video monitor. External Video Output (NTSC)Programming Input MethodTouchscreen with joystick and 3 physical rotary encoders Machine VisionReal-time Video of overlaid with stage and process variables.Tilt ControlComputer-aided 2-circle tilt control, ULTRACOLLIMATOR Measurement (option) Force Control1000grams (max) with 1 gram precision. Overall accuracy +/- 10 gramsRecipe Load & SaveUSB Port, for removable flash drive (up to 2Gb)Z Position Touch-off MethodMechanical Positioning with Force-feedback (option) sensors.X & Y PositionLASER TargetingPower Consumption300 Watts Maximum in usePower RequirementsUniversal: 100-120VAC;  200-240VACFootprint19 inches (480mm) Width x 25 inches (635 mm) Depth x 22 Inches (560mm) Height

Tool Overview

InstructionsGraphics

Operator Controls

  1. Floating polishing head assembly.
  2. Floating head lock knob.
  3. USB Slot for storing recipes
  4. Real time camera view
  5. Operator input screen. (Touch Sensitive)
  6. Emergency OFF button
  7. XYZ control joystick
  8. Dual purpose input knob
    1. X axis manual/micro move
    2. Manual table speed adjust
  9. Dual purpose input knob
    1. Y axis manual/micro move
    2. Manual tool speed adjust
  10. Z axis micro move
  11. Power button
  12. spindle shown with tool inserted

Image Removed

Operator Input Screens

InstructionsGraphics

Screen Title: ASAP - 1 Initialization Screen

Menu Label: NA

  • Displays during initial power up routine. Touch screen with stylus to exit.
Image Removed

Screen Title: Menu

Menu Label: NA

  • To access this menu, press Image Removed located in the upper left corner of
    any operator screen.

Image Removed

Screen Title: Load/Save Recipes

Menu Label: Load-Save and Tool Dia

  • This screen loads after power ON initialization screen is touched.
  • This displays machine default settings loaded on power up.
  • This screen displays current machine settings after changes.

Image Removed

Screen Title: Load-Save and Tool Diameter Select

Menu Label: Not directly accessible from the menu

  • To access this screen, first navigate to Load/Save Recipes screen shown
    above. Then click Image Removed.
  • This screen is used only to select tool diameter.

Image Removed

Screen Title: Set Scan Start/Endpoints

Menu Label: Start/Endpoints

  • This screen is primarily used to define the area to be thinned/polished.
  • Image Removed is essentially the same as pressing Run on the Run screen
  • Image Removed allows the user to interrupt the system while it is running,
    and will allow table moves and resetting of Start/End points.
  • Image Removed are temporary coordinate holders. You can store X,Y
    positions of interest to return to later.

Image Removed

Screen Title: 3D

Menu Label: 3D/Tilt

  • This screen is used to manage auto tilt and curvature parameters.

Image Removed

Screen Title:

Menu Label: Illumination/Cam Mode/Text

  • Text overlay contrast on real time camera view can be adjusted.
  • Control of camera mode is achieved on this screen.
  • This screen is used to manage power to the laser indicator.
  • This screen also controls the display of the text overlay on the
    real time camera display

Image Removed

Screen Title: Table Travel Pattern Select

Menu Label: Patterns

  • Pattern selection defines the motion of the polishing tool over the area to
    be thinned/polished.

Image Removed

Screen Title: Run

Menu Label: Run Process

  • This screen is used to control the process start and stop.
  • Checking the Timer box will force the process to stop when the timer has
    reached 0 remaining seconds.
  • Checking the Trip box will ...?
  • Checking the Zstop box will ....?
  • Vac En is not used on this tool

Image Removed

Screen Title:

Menu Label: Z Force Feedback

Image Removed

Screen Title: Keypad

Menu Label: Keypad

  • This screen is used to enter numerical values directly into parameters
  • Can be used to move the stage to absolute coordinates
    • Press Image Removed, then enter coordinate, then press
      axis label Image Removed / Image Removed / Image Removed
  • Can be used to move the stage using reletive coordinates
      • Press Image Removed, then enter coordinate, then press
        axis label Image Removed / Image Removed / Image Removed

Image Removed

Safety Precautions

This device produces visible laser radiation. To prevent injury to your eyes, never look directly into the laser.

Sample Requirements and Preparation

Expand

The Automated Selected Area Polisher (ASAP-1-IPS) is designed to thin and polish small selected areas (windows) in electronic packages and wafer-level dies. The resultant thinned component is transparent to infra-red light, thus enabling backside failure analysis to be carried out on the (front side) circuitry in the device.

Typical Applications

Image Removed

Image Removed

Image Removed

Mechanical Decapuslation

Wafer ThinningFailure AnalysisImages from http://www.ultratecusa.com

Standard Operating Procedure

Expand

Turn Power ON

ExpandInstructionsGraphics

Generally, turning the machine on only requires pressing the Green Power
button on the lower right of the control panel.

  • The Green button will light up
  • The input screen will power on
  • The real time camera view will power on
  • The illumination light will power on
  • The polishing head will move to the top of travel
  • After the tool quits moving and "Wait for RESET Initialization" is no
    longer shown on the real time camera view, use the stylus to touch
    anywhere on the input screen.
  • The input screen now displays the Load/Save Recipes screen and is
    ready for normal use.
  • The tool requires three things for power ON. Two of which are
    normally already in the correct state.
    • Toggle switch on back of tool in the down position
    • Emergency button on front panel released
    • Green power button engaged
Image Removed

Image RemovedImage Removed

Sample mounting

Expand

Your sample needs to be mounted flat and rigid to the sample holder.  Assuming mounting technique meets these two criteria, there is no preferred method.  Below are a couple common mounting techniques used regularly.

  1. Bond small samples to glass slide using Crystal Bond 555-HMP, then use clamps to hold slide in place.
    1. Place glass slide on hot plate and heat to 80 degrees C.
    2. Spread a small amount of Crystal Bond 555-HMP over the area where sample will be mounted.
    3. Press sample firmly down onto Crystal Bond and allow assembly to cool to room temperature.
    4. Place assembly onto sample holder and use mechanical clamps to hold the glass slide in place.
  2. Bond sample directly to the sample holder using Crystal Bond 555-HMP.
    1. Remove the sample holder from the ASAP-1 tool.
    2. Place on hot plate and heat to 80 degrees C.
    3. Spread a small amount of Crystal Bond 555-HMP over the area where sample will be mounted.
    4. Press sample firmly down onto Crystal Bond and allow assembly to cool to room temperature.
    5. Mount sample holder back onto the ASAP-1 tool.
  3. Mechanical Clamps
    1. Using the threaded holes provided in the sample holder, mount clamps in place as needed.  Two options on each side, one closer to the middle, the second further away from center accomodating a wide range of sample widths.
    2. The screw that threads into the clamp is used to adjust the clamp for your sample thickness.  That screw needs to be adjusted so that it sticks our futher than the sample is thick resulting the the front edge of the clamp making contact with the sample.
    3. Tighten the slotted screw until firm.
      1. Be sure to adjust the rear screw on the clamp appropriately based on the sample thickness.  You need to have the edge exerting force on the sample top, not exerting clamping force on the glass slide / sample edge.

      2. Be careful not to exert too much force on the sample.  Clamping introduces stress on the package/wafer that can result in cracking or warping during the thinning process.

Image Removed

_Proper contact angle in this image is exagerated for illustrative purposes.

Typical Sample Mounting Technique Examples

Image Removed

Image Removed

Image Removed

Wax Mount to Glass Slide and Mechanically Clamped

Mechanical Clamping for Decapsulation

Image from http://www.ultratecusa.com

Wax Mounting Direct to Sample Holder

Image from http://www.ultratecusa.com

Selected Area Polish of Wafer Backside Process Example

Expand

Set Processing Parameters

Expand

The parameters needed for backside polishing are directly observable and will not require cross sectioning of a sample as is required with decapuslation.  The tool can only polish a rectangular area with radiused corners, no other geometry is possible.  

  • Tool rotational speed (2000 rpm recommended starting point)
    1. Minimum: 1600 rpm
    2. Maximum: 2400 rpm
  • Table speed (3.5 mm/sec recommended starting point)
  • Geometry
    1. Start Point (μm)
      1. Xstart = - (Length of Area to polish in X axis-Tool Diameter) / 2
      2. Ystart = (Length of Area to polish in Y axis-Tool Diameter) / 2
    2. End Point (μm)
      1. Xend = (Length of Area to polish in X axis-Tool Diameter) / 2
      2. Yend = - (Length of Area to polish in Y axis-Tool Diameter) / 2
    3. Depth of polish in Z axis (μm)
  • Processing times
  • As a general rule, the more area to be thinned and polished the longer each step of the process will take, increase processing time accordingly.
  • Image Removed

    Select appropriate tool size and Polishing Force

    Expand

    Tools can be purchased from the manufacturer at the following link:

    https://www.ultratecusa.com/product/sap-tools/

    Area to be Polished

    * Length x Width

    (mm2)

    Tool Diameter

    (mm)

    Suggested Polishing Force

    ( grams)

    Processing factor

    (seconds/mm2)

    up to 40.4

    5

    3004 to 90.7151009 to 361.0302536 to 812.01001281 to 2253.03004.0225 to 6255.05001.5625 and up8.0

    900

    0.5
    • Corner radius need not be taken into consideration for calculating area to be polished.
    • To estimate processing time in seconds, based on the tool diameter selected, select
      the appropriate processing factor and multiply by processing area. Values are typically
      between 5 and 15 minutes.
    InstructionsGraphics

    Step 1 - Set the processing timer

    • Navigate to Keypad Screen
    • Press Image Removed, enter appropriate time in seconds then press Image Removed.

    Image Removed

    Step 2 - Set the table rastering speed (3.5 mm/s recommended)

    • Navigate to Keypad Screen
    • Click Image Removed, enter desired table speed into the field, then click Image Removed.

    Image Removed

    Step 3 - Select polishing pattern

    • Navigate to Table Travel Pattern Select screen
    • Click on the radio button below the desired pattern.
      (ASAP-1 or ASAP-1 X recommended)

    Image Removed

    Step 4 - Set polishing force

    • Navigate to Keypad Screen
    • Press Image Removed, enter appropriate force in grams, then press Image Removed.

    Image Removed

    Step 5 - Set tool diameter

    • Navigate to Load/Save Recipes screen and press Image Removed button.
    • Click radio button next to selected tool diameter

    Image Removed

    Step 6 - Verify above settings are active

    • Navigate to the Load/Save Recipes Screen
    • In the left column review:
      • Time setting
      • Speed setting
      • Selected pattern
      • Force setting
    • In the right column review:
      • Tool Dia: setting

    Image Removed

    Processing - Four Steps

    Expand
    titleClick here to expand Process Overview...

    This process recipe should be considered a guide, the procedure can be modified to meet your specific needs.
    The following procedure should yield a thinned surface with no scratches in a cavity the size of your choosing.

    Tool TypeApplicationCoolant/Slurry

    Process Time

    (minutes)

    Stage 1Fine DiamondSilicon ThinningExtender Fluid

    10 per 50 μm

    Stage 2XYLEMPre-polishBlue diamond paste and extender fluid10Stage 3XYLEMIntermediate polishYellow diamond paste and extender fluid10Stage 4XYBOVEFinal polishColloidal Silica5* Process times are flexible and you may need to run the polishing steps longer to achieve your desired results. Expand
    titleClick here to expand Processing Stage 1 - Silicon Thinning

    Processing Stage 1 - Silicon Thinning

    InstructionsGraphics

    Step 1 - Mount the selected diameter Fine Diamond tool into the spindle.

    • Make sure to align the flat on the tool shank with the set screw
      when inserting the tool into the spindle.
      Image Removed

    Image Removed

    Step 2 - Mount the sample to be thinned.

    • Wafer piece mounted backside up onto a glass slide using
      Crystal Bond 555-HMP and then mechanically clamped to the
      sample holder shown.

    Image Removed

    Step 3 - Apply one to two drops of Extender Fluid to the area to be polished.

    Image Removed

    Step 4 - Gently lower the head into operating position.
    Image Removed
    • Be careful, damageto the tool will occur, and you may be
      held responsible for damages, if you release the locking
      mechanism and allow the polishing head to drop onto the force
      sensor.

    Image Removed

    Step 5 - Define the Start point of the area to be polished.

    • Navigate to the Set Scan Start/Endpoints screen.
    • Make sure the lower right button is Image Removed indicating the
      joystick will move the spindle in the Z axis.
      • Image Removed Z axis will move very slowly
      • Image Removed Z axis will move quickly
    • Using the joystick, bring the head down until the tool nearly conacts
      the surface of your sample. Force reading should still be ~0.
      Image Removed
    • Press Image Removed button to change to Image Removed indicating the
      joystick will move the table in the X and Y axis.
    • Use the joystick to move the spindle to the upper left corner of area
      to be polished.
      Image Removed
    • Press Image Removed, then Image Removed to define the start point of the
      area to be polished.

    Image Removed

    Step 6 - Define the End point of the area to be polished.

      • Use the joystick to move the spindle to the lower right corner of area
        to be polished.
        Image Removed
      • Press Image Removed, then Image Removed to define the end point of the
        area to be polished.

    Image Removed

    Step 7 - Define the center of area to be polished as XY origin (0,0).

    • Press Image Removed to move the spindle to the center of the start and
      end points previously defined.
      • If Image Removed is displayed rather than Image Removedon the
        Set Scan Start/Endpoints screen, press Image Removed and Image Removed will
        change to Image Removed.
    • Press Image Removed, then Image Removed to define the center of the
      area to be polished. You should see the X and Y display values
      change to 0.

    Image Removed

    Image Removed

    Step 8 - Define the top surface of the sample as the origin, or zero,
    point for the Z axis.

    • Make sure force sensor is on. Image Removed
    • Make sure Image Removed is displayed before continuing.If Image Removed
      is not displayed, click Image Removed to switch joystick control to the
      Z axis.
    • Using the joystick bring the tool down to the sample surface until
      100 gm is reached. Z motion should stop automatically at 100 gm.
      Displayed on real time camera view.
    • Click Image Removed to navigate to the Keypad Screen.
    • Click Image Removed and then Image Removed. The display on real time camera
      view will update to show Z position as 0.
    • Click Image Removed, enter 20 into field, then click Image Removed. The head should
      lift 20 μm off the sample and force should read ~0. This will allow you
      to safely start the 5 point tilt auto generate process.

    Image Removed

    Image Removed

    Step 9 - Compensate for sample tilt

    • Navigate to the Set Scan Start/End screen
    • Press Image Removed to change Image Removed to Image Removed
    • Press Image Removed and the tool will proceed automatically touching
      off at 5 points within the selected area to polish.
    • Once Image Removed completes the tool will move to XY 0 and Z at 5 mm
      above the surface.
    • X and Y tilt values are shown on the real time camera view.
    • Navigate to the 3D/Tilt screen.
    • Press Image Removed followed by Image Removed. The motorized tilt stage will move
      and auto align the stage based on tilt measurement just taken.
    • Press Image Removed followed by Image Removed to measure sample tilt. The tool will
      proceed through the auto tilt measurement process as before.
    • Once auto tilt measurement completes, X and Y tilt values are shown on
      the real time camera view. The goal is to have these values measure
      at ~0. You may need to repeat this many times to achieve 0 values
      in both axis.
    • If you need to repeat, press Image Removed followed by Image Removed. The
      motorized tilt stage will move and auto align the stage based on tilt
      measurement just taken.
    • Press Image Removed followed by Image Removed to measure sample tilt and
      verify alignment.

    Image Removed

    Anchorstep10step10Step 10 - Start the spindle (2000 RPM recommended)
    • From the Keypad screen, click Image Removed, enter desired rotational speed into
      the field, then click Image Removed. The spindle will begin rotating at the
      set speed.

    Image Removed

    Step 11 - Begin the thinning process

    • Navigate to the Run screen
    • You should see your entered timer value in the Timer section
      under Set(Sec).
    • Make sure the Timer box is checked, indicating the tool should stop when
      the Elapsed time equals Set time.
    • Click Image Removed and the tool will start rastering across the selected
      area.
    • Click Image Removedto navigate to the Keypad Screen.
    • Click Image Removed, enter desired depth into field as a negative value,
      then click Image Removed. It is recommended to remove a maximum of 50 μm
      per processing step.
    • While continuing to raster, the head will slowly move down onto the part
      until the set polishing force is reached. Initially, the tool will try to maintian
      this set force by moving down in Z. Once Z has reached set point, the
      tool will hold Z position and continue processing while the applied force
      fades to zero.
    • The process will stop automatically and position the tool at the end point
      with Z ~50 μm above final Z depth when the set time has past.

    Image Removed

    Image Removed
    Image Removed

    Step 12 - Continuing to thin the sample

    • Since our extender fluid will become saturated with ground material,
      if you need to thin your sample by more than 50 μm, you should do so in
      multiple steps.
    • From the Keypad screen click Image Removed, enter 20 into field,
      then click Image Removed. The head should lift 20 μm off the sample and
      force should read ~0.
    • Use your hand to gently lift the head to the top of travel. Engage the
      locking handle by rotating the handle clockwise until tight. Make sure
      the head is locked before removing the support of your hand.
    • Wipe up the used extender fluid using a clean wipe and isopropyl
      alcohol.
    • Apply another drop or two of fresh extender fluid.
    • Gently lower the head into operating position.

    Image Removed

      • Be careful, if you release the locking mechanism and allow the
        polishing head to drop onto the force sensor, damage to the tool
        will occur.
    • Return to Step 10 above and continue. Repeat steps 10 through 12 until
      the desired thinning has been achieved.

    Rasing the head

    Image Removed

    Lowering the head

    Image Removed

    Expand
    titleClick here to expand Processing Stage 2 - Pre-polish

    Processing Stage 2 - Pre-polish

    InstructionsGraphicsStep 1 - Raise the head to allow access to the sample and tool.

    Image Removed

    Step 2 - Thoroughly clean your sample and realign for pre-polish

    • After the thinning process it is necessary to thoroughly clean your
      sample to remove large particles that would cause scratches that can't be
      removed during the pre-polish step.
    • If the sample was removed for cleaning, you must realign and clamp your
      sample, see these instructions.

    Step 3 - Mount the selected diameter Xylem tool into the spindle.

    • Make sure to mount the tool reserved for the blue diamond paste, it will
      be marked with blue paint on the mounting flat.
    • Make sure to align the flat on the tool shank with the set screw
      when inserting the tool into the spindle.

    Image Removed

    Step 4 - Apply a small amount of Blue Diamond paste to the thinned area.

    Image Removed

    Step 5 - Apply one to two drops of Extender Fluid to the thinned area
    and spread blue diamond paste.

    Image Removed

    Step 6 - Gently lower the head into operating position.

    Image Removed
      • Be careful, if you release the locking mechanism and allow the
        polishing head to drop onto the force sensor, damage to the tool
        will occur.

    Image Removed

    Step 7 - Define the top surface of the sample as the origin, or zero,
    point for the Z axis.

    • Since you have changed tools, we must re establish the Z origin.
    • Make sure force sensor is on. Image Removed
    • Make sure Image Removed is displayed before continuing. If Image Removed
      is not displayed, click Image Removed to switch joystick control to the
      Z axis.
    • Using the joystick bring the tool down to the sample surface until
      100 gm is reached. Z motion should stop automatically at 100 gm.
      Displayed on real time camera view.
    • Click Image Removed to navigate to the Keypad Screen.
    • Click Image Removed and then Image Removed. The display on real time camera
      view will update to show Z position as 0.
    • Click Image Removed, enter 50 into field, then click Image Removed. The head will
      lift 50 μm off the sample and force should read ~0.

    Image Removed

    Image Removed

    Image Removed

    Step 8 - Move to the center of the selected area

    • Navigate to the Set Scan Start/Endpoints screen
    • Press Image Removed. The head will move to the defined XY origin.

    Step 9 - Change the timer for the polishing process if applicable. If no
    change needed, skip to next step.
    (600 seconds in this example)

    • Navigate to the Keypad screen.
    • Click Image Removed, enter processing time into the field, then click Image Removed.
      This will not produce an immediate and noticable result, however it does
      set the Run timer shown on the Run screen.
    • Follow this step's instructions to reset the timer. Setting the timer,
      even to the same value, resets the timer. However, the displayed elapse
      time will not change to 0 until the Image Removed button is pressed on
      the Run screen.

    Image Removed

    Step 10 - Start the spindle (2000 RPM recommended)

    • From the Keypad screen, click Image Removed, enter desired rotational speed into
      the field, then click Image Removed. The spindle will begin rotating at the
      set speed.

    Image Removed

    Step 11 - Begin the pre-polishing process

    • Navigate to the Run screen
    • You should see your entered timer value in the Timer section
      under Set(Sec).
    • Make sure the Timer box is checked, indicating the tool should stop when
      the Elapsed time equals Set time.
    • Click Image Removed and the tool will start rastering across the selected
      area.
    • Click Image Removedto navigate to the Keypad Screen.
    • Click Image Removed, enter an unreachable depth into field as a negative value,
      then click Image Removed. This ensures you are controlled by the timer.
    • While continuing to raster, the head will slowly move down onto the part
      until the set polishing force is reached.
    • The process will stop automatically and position the tool at the end point
      with Z ~50 μm above final Z depth when the set time has past.

    Image Removed

    Image Removed

    Image Removed

    Expand
    titleClick here to expand Processing Stage 3 - Intermediate polish

    Processing Stage 3 - Intermediate polish

    InstructionsGraphicsStep 1 - Raise the head to allow access to the sample and tool.

    Image Removed

    Step 2 - Thoroughly clean your sample and realign for intermediate polish

    • After the blue compound it is necessary to thoroughly clean your
      sample to remove large particles that would cause scratches that can't be
      removed during the intermediate polish step.
    • If the sample was removed for cleaning, you must realign and clamp your
      sample, see these instructions.

    Step 3 - Mount the selected diameter Xylem tool into the spindle.

    • Make sure to mount the tool reserved for the yellow diamond paste, it will
      be marked with yellow paint on the mounting flat.
    • Make sure to align the flat on the tool shank with the set screw
      when inserting the tool into the spindle.

    Image Removed

    Step 4 - Apply a small amount of Yellow Diamond paste to the thinned area.

    Image Removed

    Step 5 - Apply one to two drops of Extender Fluid to the thinned area
    and spread yellow diamond paste.

    Image Removed

    Step 6 - Gently lower the head into operating position.

    Image Removed
      • Be careful, if you release the locking mechanism and allow the
        polishing head to drop onto the force sensor, damage to the tool
        will occur.

    Image Removed

    Step 7 - Define the top surface of the sample as the origin, or zero,
    point for the Z axis.

    • Since you have changed tools, we must re establish the Z origin.
    • Make sure force sensor is on. Image Removed
    • Make sure Image Removed is displayed before continuing. If Image Removed
      is not displayed, click Image Removed to switch joystick control to the
      Z axis.
    • Using the joystick bring the tool down to the sample surface until
      100 gm is reached. Z motion should stop automatically at 100 gm.
      Displayed on real time camera view.
    • Click Image Removed to navigate to the Keypad Screen.
    • Click Image Removed and then Image Removed. The display on real time camera
      view will update to show Z position as 0.
    • Click Image Removed, enter 50 into field, then click Image Removed. The head will
      lift 50 μm off the sample and force should read ~0.

    Image Removed

    Image Removed

    Image Removed

    Step 8 - Move to the center of the selected area

    • Navigate to the Set Scan Start/Endpoints screen
    • Press Image Removed. The head will move to the defined XY origin.

    Step 9 - Change the timer for the polishing process if applicable. If no
    change needed, skip to next step.
    (600 seconds in this example)

    • Navigate to the Keypad screen.
    • Click Image Removed, enter processing time into the field, then click Image Removed.
      This will not produce an immediate and noticeable result, however it does
      set the Run timer shown on the Run screen.
    • Follow this step's instructions to reset the timer. Setting the timer,
      even to the same value, resets the timer. However, the displayed elapse
      time will not change to 0 until the Image Removed button is pressed on
      the Run screen.

    Image Removed

    Step 10 - Start the spindle (2000 RPM recommended)

    • From the Keypad screen, click Image Removed, enter desired rotational speed into
      the field, then click Image Removed. The spindle will begin rotating at the
      set speed.

    Image Removed

    Step 11 - Begin the intermediate polishing process

    • Navigate to the Run screen
    • You should see your entered timer value in the Timer section
      under Set(Sec).
    • Make sure the Timer box is checked, indicating the tool should stop when
      the Elapsed time equals Set time.
    • Click Image Removed and the tool will start rastering across the selected
      area.
    • Click Image Removedto navigate to the Keypad Screen.
    • Click Image Removed, enter an unreachable depth into field as a negative value,
      then click Image Removed. This ensures you are controlled by the timer.
    • While continuing to raster, the head will slowly move down onto the part
      until the set polishing force is reached.
    • The process will stop automatically and position the tool at the end point
      with Z ~50 μm above final Z depth when the set time has past.

    Image Removed

    Image Removed

    Image Removed

    Expand
    titleClick here to expand Processing Stage 4 - Final polish

    Processing Stage 4 - Final polish

    InstructionsGraphicsStep 1 - Raise the head to allow access to the sample and tool.

    Image Removed

    Step 2 - Thoroughly clean your sample and realign for intermediate polish

    • After the blue compound it is necessary to thoroughly clean your
      sample to remove large particles that would cause scratches that can't be
      removed during the intermediate polish step.
    • If the sample was removed for cleaning, you must realign and clamp your
      sample, see these instructions.

    Step 3 - Mount the selected diameter Xbove tool into the spindle.

    • Make sure to mount the tool reserved for the yellow diamond paste, it will
      be marked with yellow paint on the mounting flat.
    • Make sure to align the flat on the tool shank with the set screw
      when inserting the tool into the spindle.

    Image Removed

    Step 4 - Apply one to two drops of Extender Fluid to the thinned area
    and spread yellow diamond paste.

    Image Removed

    Step 5 - Gently lower the head into operating position.

    Image Removed
      • Be careful, if you release the locking mechanism and allow the
        polishing head to drop onto the force sensor, damage to the tool
        will occur.

    Image Removed

    Step 6 - Define the top surface of the sample as the origin, or zero,
    point for the Z axis.

    • Since you have changed tools, we must re establish the Z origin.
    • Make sure force sensor is on. Image Removed
    • Make sure Image Removed is displayed before continuing. If Image Removed
      is not displayed, click Image Removed to switch joystick control to the
      Z axis.
    • Using the joystick bring the tool down to the sample surface until
      100 gm is reached. Z motion should stop automatically at 100 gm.
      Displayed on real time camera view.
    • Click Image Removed to navigate to the Keypad Screen.
    • Click Image Removed and then Image Removed. The display on real time camera
      view will update to show Z position as 0.
    • Click Image Removed, enter 50 into field, then click Image Removed. The head will
      lift 50 μm off the sample and force should read ~0.

    Image Removed

    Image Removed

    Image Removed

    Step 7 - Move to the center of the selected area

    • Navigate to the Set Scan Start/Endpoints screen
    • Press Image Removed. The head will move to the defined XY origin.

    Step 8 - Change the timer for the polishing process if applicable. If no
    change needed, skip to next step.
    (600 seconds in this example)

    • Navigate to the Keypad screen.
    • Click Image Removed, enter processing time into the field, then click Image Removed.
      This will not produce an immediate and noticeable result, however it does
      set the Run timer shown on the Run screen.
    • Follow this step's instructions to reset the timer. Setting the timer,
      even to the same value, resets the timer. However, the displayed elapse
      time will not change to 0 until the Image Removed button is pressed on
      the Run screen.

    Image Removed

    Step 9 - Start the spindle (2000 RPM recommended)

    • From the Keypad screen, click Image Removed, enter desired rotational speed into
      the field, then click Image Removed. The spindle will begin rotating at the
      set speed.

    Image Removed

    Step 10 - Begin the Final polishing process

    • Navigate to the Run screen
    • You should see your entered timer value in the Timer section
      under Set(Sec).
    • Make sure the Timer box is checked, indicating the tool should stop when
      the Elapsed time equals Set time.
    • Click Image Removed and the tool will start rastering across the selected
      area.
    • Click Image Removedto navigate to the Keypad Screen.
    • Click Image Removed, enter an unreachable depth into field as a negative value,
      then click Image Removed. This ensures you are controlled by the timer.
    • While continuing to raster, the head will slowly move down onto the part
      until the set polishing force is reached.
    • The process will stop automatically and position the tool at the end point
      with Z ~50 μm above final Z depth when the set time has past.

    Image Removed

    Image Removed

    Image Removed

    Questions & Troubleshooting

    Process Library

    References

    Manufacturer Website

    ASAP-1 IPS Data Sheet

    ASAP-1 Product Brochure

    ASAP-1IPS Rev1.74.pdf

    ASAP-1IPS Rev1.76.pdf

    SAP presentation.pptx

    processing time calc.xlsx

    Widget Connector
    urlhttps://www.youtube.com/watch?v=O7rpZgH0y9Q

    Widget Connector
    urlhttps://www.youtube.com/watch?v=F3cyXxYKPjM

    Widget Connector
    urlhttps://www.youtube.com/watch?v=wvxIs3-qAgs

    Widget Connector
    urlhttps://www.youtube.com/watch?v=mVqJLc3-_e8

    Widget Connector
    urlhttps://www.youtube.com/watch?v=LMbsIjKutNQ

    Mihailo Bradash
    Location:
     BRK 2221
    Maximum Wafer Size: 
    4"/100 mm


    Table of Content Zone
    locationtop
    styledisc

    Table of Contents
    outlinetrue
    indent25px
    stylenone


    Overview

    Expand

    Note: the following is quoted from ULTRA TEC's documentation

    General Description

    ASAP-1® IPS is a digital sample preparation system for the decapsulation, thinning and polish­ing of packaged wafer-level devices.

    Drawing on ULTRA TEC’s market knowledge and long-standing leadership in electronic sample preparation, IPS is ‘device centric’ – designed to meet the non-destructive, high yield and survivability needs of micro-electronic components.

    IPS’ Touch Screen OS has been designed from the ground-up to control all aspects of the sample preparation process. Advanced programming drives IPS’s deep sub-micron axes of motion. The machine vision monitor provides an unique ‘always-live’ image of the device, along with key navigational and process information.

    The touch-off tool control provides Z-position at pre-defined locations and pressures – enabling package and die tilt to be corrected quickly and accurately. Tool height variations are automatically-calibrated.

    Selected area preparation

    "Selected Area Preparation (SAP) has long been established as a low damage machining method, with the use of relatively low spindle speeds, in conjunction with the floating head providing very low temperature rises. This is in stark contrast with high speed 'mills' which introduce excess vibration, ultrasonics, and process temperature rises -- generally requiring either a cooling bath to be in position at all time, or even a constant high flow of coolant to avoid device overheating. SAP is almost always a ‘wet process’ . However the coolant required is minimal."

    The use of ULTRA TEC’s patented 'float down' head approach has enabled several hundred customers to achieve high quality polished surfaces, ready for analysis. High yields have been shown using analog ASAP-1® equipment for thinning silicon to less than 20 micron target RST. The move to higher accuracy digital systems allows for routine thinning of extremely thin substrates at even low single-digit RST’s. The move to larger flip chips mounted on PCB substrates requires inherent design for 3D sample prep. Sample Preparation developments have been driven by the industry's push for both smaller RST’s – for optimal backside imaging and attenuation – and the need for improved uniformity, to satisfy the latest analytical techniques.

    • “A Straightforward Guide to the Sample Preparation of Curved & Warped IC's”; TEC Note #16, 2015 • “Backside Failure Analysis Techniques: What’s The Gain Of Silicon Getting Thinner”; Boit, Schafer et al; ESREF 2014 • “The State of the Art in Backside Sample Prep”; Jim Colvin, EDFA Vol 4, No.2, 2002 • “Packages Have Become the New IC’s”; Jim Colvin, EDFA Volume 16, Issue 4, November 2014

    Specifications

    • Suits all sizes of die - package, wafer and board-level
    • Real Time Video Monitor with system parameter
    • Touchscreen control with physical joystick & controls
    • Rigidized Table Assembly and closed-loop, high-torque, motor control enhances the machining of tough and hard materials
    • X, Y and Z axes all have deep sub-micron accuracy
    • Accurately decaps, then thins substrate and polishes
    • Patented Floating Head provides a true polishing action -- yields polishing quality & high survivability
    • Intuitive menus provide a powerful, easy to use, system
    • USB Flash Driveinterface for preparation recipe storage
    • Short set-up and process times
    • Accurate die-tilt adjustment ‘on the fly’
    • Bench-top & Quiet in Operation
    Z-Vertical Direction Precision0.04 microns (40 nanometers)
    Table Precision (X& Y Travel)0.2 microns (200 nanometers)
    Table Travel Amplitude100mm x 100mm
    Polishing MethodPatented ASAP-1 Float-down head, with Z-lock, enhanced with electronic sensors and tool patterns
    VideoReal-time machine vision with 6.5 inch video monitor. External Video Output (NTSC)
    Programming Input MethodTouchscreen with joystick and 3 physical rotary encoders 
    Machine VisionReal-time Video of overlaid with stage and process variables.
    Tilt ControlComputer-aided 2-circle tilt control, ULTRACOLLIMATOR Measurement (option) 
    Force Control1000grams (max) with 1 gram precision. Overall accuracy +/- 10 grams
    Recipe Load & SaveUSB Port, for removable flash drive (up to 2Gb)
    Z Position Touch-off MethodMechanical Positioning with Force-feedback (option) sensors.
    X & Y PositionLASER Targeting
    Power Consumption300 Watts Maximum in use
    Power RequirementsUniversal: 100-120VAC;  200-240VAC
    Footprint19 inches (480mm) Width x 25 inches (635 mm) Depth x 22 Inches (560mm) Height

    Tool Overview

    InstructionsGraphics

    Operator Controls

    1. Floating polishing head assembly.
    2. Floating head lock knob.
    3. USB Slot for storing recipes
    4. Real time camera view
    5. Operator input screen. (Touch Sensitive)
    6. Emergency OFF button
    7. XYZ control joystick
    8. Dual purpose input knob
      1. X axis manual/micro move
      2. Manual table speed adjust
    9. Dual purpose input knob
      1. Y axis manual/micro move
      2. Manual tool speed adjust
    10. Z axis micro move
    11. Power button
    12. spindle shown with tool inserted


    Operator Input Screens

    InstructionsGraphics

    Screen Title: ASAP - 1 Initialization Screen

    Menu Label: NA

    • Displays during initial power up routine. Touch screen with stylus to exit.


    Screen Title: Menu

    Menu Label: NA

    • To access this menu, press located in the upper left corner of
      any operator screen.

    Screen Title: Load/Save Recipes

    Menu Label: Load-Save and Tool Dia

    • This screen loads after power ON initialization screen is touched.
    • This displays machine default settings loaded on power up.
    • This screen displays current machine settings after changes.

    Screen Title: Load-Save and Tool Diameter Select

    Menu Label: Not directly accessible from the menu

    • To access this screen, first navigate to Load/Save Recipes screen shown
      above. Then click .
    • This screen is used only to select tool diameter.

    Screen Title: Set Scan Start/Endpoints

    Menu Label: Start/Endpoints

    • This screen is primarily used to define the area to be thinned/polished.
    • is essentially the same as pressing Run on the Run screen
    • allows the user to interrupt the system while it is running,
      and will allow table moves and resetting of Start/End points.
    • are temporary coordinate holders. You can store X,Y
      positions of interest to return to later.

    Screen Title: 3D

    Menu Label: 3D/Tilt

    • This screen is used to manage auto tilt and curvature parameters.

    Screen Title:

    Menu Label: Illumination/Cam Mode/Text

    • Text overlay contrast on real time camera view can be adjusted.
    • Control of camera mode is achieved on this screen.
    • This screen is used to manage power to the laser indicator.
    • This screen also controls the display of the text overlay on the
      real time camera display

    Screen Title: Table Travel Pattern Select

    Menu Label: Patterns

    • Pattern selection defines the motion of the polishing tool over the area to
      be thinned/polished.

    Screen Title: Run

    Menu Label: Run Process

    • This screen is used to control the process start and stop.
    • Checking the Timer box will force the process to stop when the timer has
      reached 0 remaining seconds.
    • Checking the Trip box will ...?
    • Checking the Zstop box will ....?
    • Vac En is not used on this tool

    Screen Title:

    Menu Label: Z Force Feedback

    Screen Title: Keypad

    Menu Label: Keypad

    • This screen is used to enter numerical values directly into parameters
    • Can be used to move the stage to absolute coordinates
      • Press , then enter coordinate, then press
        axis label / /
    • Can be used to move the stage using reletive coordinates
        • Press , then enter coordinate, then press
          axis label / /


    Safety Precautions

    This device produces visible laser radiation. To prevent injury to your eyes, never look directly into the laser.



    Sample Requirements and Preparation

    Expand

    The Automated Selected Area Polisher (ASAP-1-IPS) is designed to thin and polish small selected areas (windows) in electronic packages and wafer-level dies. The resultant thinned component is transparent to infra-red light, thus enabling backside failure analysis to be carried out on the (front side) circuitry in the device.

    Typical Applications

    Mechanical Decapuslation

    Wafer ThinningFailure Analysis
    Images from http://www.ultratecusa.com



    Standard Operating Procedure

    Expand

    Turn Power ON

    Expand


    InstructionsGraphics

    Generally, turning the machine on only requires pressing the Green Power
    button on the lower right of the control panel.

    • The Green button will light up
    • The input screen will power on
    • The real time camera view will power on
    • The illumination light will power on
    • The polishing head will move to the top of travel
    • After the tool quits moving and "Wait for RESET Initialization" is no
      longer shown on the real time camera view, use the stylus to touch
      anywhere on the input screen.
    • The input screen now displays the Load/Save Recipes screen and is
      ready for normal use.
    • The tool requires three things for power ON. Two of which are
      normally already in the correct state.
      • Toggle switch on back of tool in the down position
      • Emergency button on front panel released
      • Green power button engaged



    Sample mounting

    Expand

    Your sample needs to be mounted flat and rigid to the sample holder.  Assuming mounting technique meets these two criteria, there is no preferred method.  Below are a couple common mounting techniques used regularly.

    1. Bond small samples to glass slide using Crystal Bond 555-HMP, then use clamps to hold slide in place.
      1. Place glass slide on hot plate and heat to 80 degrees C.
      2. Spread a small amount of Crystal Bond 555-HMP over the area where sample will be mounted.
      3. Press sample firmly down onto Crystal Bond and allow assembly to cool to room temperature.
      4. Place assembly onto sample holder and use mechanical clamps to hold the glass slide in place.
    2. Bond sample directly to the sample holder using Crystal Bond 555-HMP.
      1. Remove the sample holder from the ASAP-1 tool.
      2. Place on hot plate and heat to 80 degrees C.
      3. Spread a small amount of Crystal Bond 555-HMP over the area where sample will be mounted.
      4. Press sample firmly down onto Crystal Bond and allow assembly to cool to room temperature.
      5. Mount sample holder back onto the ASAP-1 tool.
    3. Mechanical Clamps
      1. Using the threaded holes provided in the sample holder, mount clamps in place as needed.  Two options on each side, one closer to the middle, the second further away from center accomodating a wide range of sample widths.
      2. The screw that threads into the clamp is used to adjust the clamp for your sample thickness.  That screw needs to be adjusted so that it sticks our futher than the sample is thick resulting the the front edge of the clamp making contact with the sample.
      3. Tighten the slotted screw until firm.
        1. Be sure to adjust the rear screw on the clamp appropriately based on the sample thickness.  You need to have the edge exerting force on the sample top, not exerting clamping force on the glass slide / sample edge.

        2. Be careful not to exert too much force on the sample.  Clamping introduces stress on the package/wafer that can result in cracking or warping during the thinning process.

    _Proper contact angle in this image is exagerated for illustrative purposes.


    Typical Sample Mounting Technique Examples

    Wax Mount to Glass Slide and Mechanically Clamped

    Mechanical Clamping for Decapsulation

    Image from http://www.ultratecusa.com

    Wax Mounting Direct to Sample Holder

    Image from http://www.ultratecusa.com


    Selected Area Polish of Wafer Backside Process Example

    Expand

    Set Processing Parameters

    Expand

    The parameters needed for backside polishing are directly observable and will not require cross sectioning of a sample as is required with decapuslation.  The tool can only polish a rectangular area with radiused corners, no other geometry is possible.  

    1. Tool rotational speed (2000 rpm recommended starting point)
      1. Minimum: 1600 rpm
      2. Maximum: 2400 rpm
    2. Table speed (3.5 mm/sec recommended starting point)
    3. Geometry
      1. Start Point (μm)
        1. Xstart = - (Length of Area to polish in X axis-Tool Diameter) / 2
        2. Ystart = (Length of Area to polish in Y axis-Tool Diameter) / 2
      2. End Point (μm)
        1. Xend = (Length of Area to polish in X axis-Tool Diameter) / 2
        2. Yend = - (Length of Area to polish in Y axis-Tool Diameter) / 2
      3. Depth of polish in Z axis (μm)
    4. Processing times
      1. As a general rule, the more area to be thinned and polished the longer each step of the process will take, increase processing time accordingly.

    Select appropriate tool size and Polishing Force

    Expand

    Tools can be purchased from the manufacturer at the following link:

    https://www.ultratecusa.com/product/sap-tools/

    Area to be Polished

    * Length x Width

    (mm2)

    Tool Diameter

    (mm)

    Suggested Polishing Force

    ( grams)

    Processing factor

    (seconds/mm2)

    up to 40.4

    5

    300
    4 to 90.715100
    9 to 361.03025
    36 to 812.010012
    81 to 2253.03004.0
    225 to 6255.05001.5
    625 and up8.0

    900

    0.5
    • Corner radius need not be taken into consideration for calculating area to be polished.
    • To estimate processing time in seconds, based on the tool diameter selected, select
      the appropriate processing factor and multiply by processing area. Values are typically
      between 5 and 15 minutes.


    InstructionsGraphics

    Step 1 - Set the processing timer

    • Navigate to Keypad Screen
    • Press , enter appropriate time in seconds then press .

    Step 2 - Set the table rastering speed (3.5 mm/s recommended)

    • Navigate to Keypad Screen
    • Click , enter desired table speed into the field, then click .

    Step 3 - Select polishing pattern

    • Navigate to Table Travel Pattern Select screen
    • Click on the radio button below the desired pattern.
      (ASAP-1 or ASAP-1 X recommended)

    Step 4 - Set polishing force

    • Navigate to Keypad Screen
    • Press , enter appropriate force in grams, then press .

    Step 5 - Set tool diameter

    • Navigate to Load/Save Recipes screen and press button.
    • Click radio button next to selected tool diameter

    Step 6 - Verify above settings are active

    • Navigate to the Load/Save Recipes Screen
    • In the left column review:
      • Time setting
      • Speed setting
      • Selected pattern
      • Force setting
    • In the right column review:
      • Tool Dia: setting


    Processing - Four Steps

    Expand
    titleClick here to expand Process Overview...

    This process recipe should be considered a guide, the procedure can be modified to meet your specific needs.
    The following procedure should yield a thinned surface with no scratches in a cavity the size of your choosing.


    Tool TypeApplicationCoolant/Slurry

    Process Time

    (minutes)

    Stage 1Fine DiamondSilicon ThinningExtender Fluid

    10 per 50 μm

    Stage 2XYLEMPre-polishBlue diamond paste and extender fluid10
    Stage 3XYLEMIntermediate polishYellow diamond paste and extender fluid10
    Stage 4XYBOVEFinal polishColloidal Silica5
    * Process times are flexible and you may need to run the polishing steps longer to achieve your desired results.



    Expand
    titleClick here to expand Processing Stage 1 - Silicon Thinning

    Processing Stage 1 - Silicon Thinning

    InstructionsGraphics

    Step 1 - Mount the selected diameter Fine Diamond tool into the spindle.

    • Make sure to align the flat on the tool shank with the set screw
      when inserting the tool into the spindle.

    Step 2 - Mount the sample to be thinned.

    • Wafer piece mounted backside up onto a glass slide using
      Crystal Bond 555-HMP and then mechanically clamped to the
      sample holder shown.




    Step 3 - Apply one to two drops of Extender Fluid to the area to be polished.

    Step 4 - Gently lower the head into operating position.
    • Be careful, damageto the tool will occur, and you may be
      held responsible for damages, if you release the locking
      mechanism and allow the polishing head to drop onto the force
      sensor.

    Step 5 - Define the Start point of the area to be polished.

    • Navigate to the Set Scan Start/Endpoints screen.
    • Make sure the lower right button is indicating the
      joystick will move the spindle in the Z axis.
      • Z axis will move very slowly
      • Z axis will move quickly
    • Using the joystick, bring the head down until the tool nearly conacts
      the surface of your sample. Force reading should still be ~0.
    • Press button to change to indicating the
      joystick will move the table in the X and Y axis.
    • Use the joystick to move the spindle to the upper left corner of area
      to be polished.
    • Press , then to define the start point of the
      area to be polished.

    Step 6 - Define the End point of the area to be polished.

      • Use the joystick to move the spindle to the lower right corner of area
        to be polished.
      • Press , then to define the end point of the
        area to be polished.


    Step 7 - Define the center of area to be polished as XY origin (0,0).

    • Press to move the spindle to the center of the start and
      end points previously defined.
      • If is displayed rather than on the
        Set Scan Start/Endpoints screen, press and will
        change to .
    • Press , then to define the center of the
      area to be polished. You should see the X and Y display values
      change to 0.

    Step 8 - Define the top surface of the sample as the origin, or zero,
    point for the Z axis.

    • Make sure force sensor is on.
    • Make sure is displayed before continuing.If
      is not displayed, click to switch joystick control to the
      Z axis.
    • Using the joystick bring the tool down to the sample surface until
      100 gm is reached. Z motion should stop automatically at 100 gm.
      Displayed on real time camera view.
    • Click to navigate to the Keypad Screen.
    • Click and then . The display on real time camera
      view will update to show Z position as 0.
    • Click , enter 20 into field, then click . The head should
      lift 20 μm off the sample and force should read ~0. This will allow you
      to safely start the 5 point tilt auto generate process.

    Step 9 - Compensate for sample tilt

    • Navigate to the Set Scan Start/End screen
    • Press to change to
    • Press and the tool will proceed automatically touching
      off at 5 points within the selected area to polish.
    • Once completes the tool will move to XY 0 and Z at 5 mm
      above the surface.
    • X and Y tilt values are shown on the real time camera view.
    • Navigate to the 3D/Tilt screen.
    • Press followed by . The motorized tilt stage will move
      and auto align the stage based on tilt measurement just taken.
    • Press followed by to measure sample tilt. The tool will
      proceed through the auto tilt measurement process as before.
    • Once auto tilt measurement completes, X and Y tilt values are shown on
      the real time camera view. The goal is to have these values measure
      at ~0. You may need to repeat this many times to achieve 0 values
      in both axis.
    • If you need to repeat, press followed by . The
      motorized tilt stage will move and auto align the stage based on tilt
      measurement just taken.
    • Press followed by to measure sample tilt and
      verify alignment.

    Anchor
    step10
    step10
    Step 10 - Start the spindle (2000 RPM recommended)

    • From the Keypad screen, click , enter desired rotational speed into
      the field, then click . The spindle will begin rotating at the
      set speed.

    Step 11 - Begin the thinning process

    • Navigate to the Run screen
    • You should see your entered timer value in the Timer section
      under Set(Sec).
    • Make sure the Timer box is checked, indicating the tool should stop when
      the Elapsed time equals Set time.
    • Click and the tool will start rastering across the selected
      area.
    • Click to navigate to the Keypad Screen.
    • Click , enter desired depth into field as a negative value,
      then click . It is recommended to remove a maximum of 50 μm
      per processing step.
    • While continuing to raster, the head will slowly move down onto the part
      until the set polishing force is reached. Initially, the tool will try to maintian
      this set force by moving down in Z. Once Z has reached set point, the
      tool will hold Z position and continue processing while the applied force
      fades to zero.
    • The process will stop automatically and position the tool at the end point
      with Z ~50 μm above final Z depth when the set time has past.


    Step 12 - Continuing to thin the sample

    • Since our extender fluid will become saturated with ground material,
      if you need to thin your sample by more than 50 μm, you should do so in
      multiple steps.
    • From the Keypad screen click , enter 20 into field,
      then click . The head should lift 20 μm off the sample and
      force should read ~0.
    • Use your hand to gently lift the head to the top of travel. Engage the
      locking handle by rotating the handle clockwise until tight. Make sure
      the head is locked before removing the support of your hand.
    • Wipe up the used extender fluid using a clean wipe and isopropyl
      alcohol.
    • Apply another drop or two of fresh extender fluid.
    • Gently lower the head into operating position.

      • Be careful, if you release the locking mechanism and allow the
        polishing head to drop onto the force sensor, damage to the tool
        will occur.
    • Return to Step 10 above and continue. Repeat steps 10 through 12 until
      the desired thinning has been achieved.


    Rasing the head


    Lowering the head




    Expand
    titleClick here to expand Processing Stage 2 - Pre-polish

    Processing Stage 2 - Pre-polish

    InstructionsGraphics
    Step 1 - Raise the head to allow access to the sample and tool.

    Step 2 - Thoroughly clean your sample and realign for pre-polish

    • After the thinning process it is necessary to thoroughly clean your
      sample to remove large particles that would cause scratches that can't be
      removed during the pre-polish step.
    • If the sample was removed for cleaning, you must realign and clamp your
      sample, see these instructions.

    Step 3 - Mount the selected diameter Xylem tool into the spindle.

    • Make sure to mount the tool reserved for the blue diamond paste, it will
      be marked with blue paint on the mounting flat.
    • Make sure to align the flat on the tool shank with the set screw
      when inserting the tool into the spindle.

    Step 4 - Apply a small amount of Blue Diamond paste to the thinned area.


    Step 5 - Apply one to two drops of Extender Fluid to the thinned area
    and spread blue diamond paste.

    Step 6 - Gently lower the head into operating position.

      • Be careful, if you release the locking mechanism and allow the
        polishing head to drop onto the force sensor, damage to the tool
        will occur.

    Step 7 - Define the top surface of the sample as the origin, or zero,
    point for the Z axis.

    • Since you have changed tools, we must re establish the Z origin.
    • Make sure force sensor is on.
    • Make sure is displayed before continuing. If
      is not displayed, click to switch joystick control to the
      Z axis.
    • Using the joystick bring the tool down to the sample surface until
      100 gm is reached. Z motion should stop automatically at 100 gm.
      Displayed on real time camera view.
    • Click to navigate to the Keypad Screen.
    • Click and then . The display on real time camera
      view will update to show Z position as 0.
    • Click , enter 50 into field, then click . The head will
      lift 50 μm off the sample and force should read ~0.

    Step 8 - Move to the center of the selected area

    • Navigate to the Set Scan Start/Endpoints screen
    • Press . The head will move to the defined XY origin.

    Step 9 - Change the timer for the polishing process if applicable. If no
    change needed, skip to next step.
    (600 seconds in this example)

    • Navigate to the Keypad screen.
    • Click , enter processing time into the field, then click .
      This will not produce an immediate and noticable result, however it does
      set the Run timer shown on the Run screen.
    • Follow this step's instructions to reset the timer. Setting the timer,
      even to the same value, resets the timer. However, the displayed elapse
      time will not change to 0 until the button is pressed on
      the Run screen.

    Step 10 - Start the spindle (2000 RPM recommended)

    • From the Keypad screen, click , enter desired rotational speed into
      the field, then click . The spindle will begin rotating at the
      set speed.

    Step 11 - Begin the pre-polishing process

    • Navigate to the Run screen
    • You should see your entered timer value in the Timer section
      under Set(Sec).
    • Make sure the Timer box is checked, indicating the tool should stop when
      the Elapsed time equals Set time.
    • Click and the tool will start rastering across the selected
      area.
    • Click to navigate to the Keypad Screen.
    • Click , enter an unreachable depth into field as a negative value,
      then click . This ensures you are controlled by the timer.
    • While continuing to raster, the head will slowly move down onto the part
      until the set polishing force is reached.
    • The process will stop automatically and position the tool at the end point
      with Z ~50 μm above final Z depth when the set time has past.



    Expand
    titleClick here to expand Processing Stage 3 - Intermediate polish

    Processing Stage 3 - Intermediate polish

    InstructionsGraphics
    Step 1 - Raise the head to allow access to the sample and tool.

    Step 2 - Thoroughly clean your sample and realign for intermediate polish

    • After the blue compound it is necessary to thoroughly clean your
      sample to remove large particles that would cause scratches that can't be
      removed during the intermediate polish step.
    • If the sample was removed for cleaning, you must realign and clamp your
      sample, see these instructions.

    Step 3 - Mount the selected diameter Xylem tool into the spindle.

    • Make sure to mount the tool reserved for the yellow diamond paste, it will
      be marked with yellow paint on the mounting flat.
    • Make sure to align the flat on the tool shank with the set screw
      when inserting the tool into the spindle.

    Step 4 - Apply a small amount of Yellow Diamond paste to the thinned area.

    Step 5 - Apply one to two drops of Extender Fluid to the thinned area
    and spread yellow diamond paste.

    Step 6 - Gently lower the head into operating position.

      • Be careful, if you release the locking mechanism and allow the
        polishing head to drop onto the force sensor, damage to the tool
        will occur.

    Step 7 - Define the top surface of the sample as the origin, or zero,
    point for the Z axis.

    • Since you have changed tools, we must re establish the Z origin.
    • Make sure force sensor is on.
    • Make sure is displayed before continuing. If
      is not displayed, click to switch joystick control to the
      Z axis.
    • Using the joystick bring the tool down to the sample surface until
      100 gm is reached. Z motion should stop automatically at 100 gm.
      Displayed on real time camera view.
    • Click to navigate to the Keypad Screen.
    • Click and then . The display on real time camera
      view will update to show Z position as 0.
    • Click , enter 50 into field, then click . The head will
      lift 50 μm off the sample and force should read ~0.

    Step 8 - Move to the center of the selected area

    • Navigate to the Set Scan Start/Endpoints screen
    • Press . The head will move to the defined XY origin.

    Step 9 - Change the timer for the polishing process if applicable. If no
    change needed, skip to next step.
    (600 seconds in this example)

    • Navigate to the Keypad screen.
    • Click , enter processing time into the field, then click .
      This will not produce an immediate and noticeable result, however it does
      set the Run timer shown on the Run screen.
    • Follow this step's instructions to reset the timer. Setting the timer,
      even to the same value, resets the timer. However, the displayed elapse
      time will not change to 0 until the button is pressed on
      the Run screen.

    Step 10 - Start the spindle (2000 RPM recommended)

    • From the Keypad screen, click , enter desired rotational speed into
      the field, then click . The spindle will begin rotating at the
      set speed.

    Step 11 - Begin the intermediate polishing process

    • Navigate to the Run screen
    • You should see your entered timer value in the Timer section
      under Set(Sec).
    • Make sure the Timer box is checked, indicating the tool should stop when
      the Elapsed time equals Set time.
    • Click and the tool will start rastering across the selected
      area.
    • Click to navigate to the Keypad Screen.
    • Click , enter an unreachable depth into field as a negative value,
      then click . This ensures you are controlled by the timer.
    • While continuing to raster, the head will slowly move down onto the part
      until the set polishing force is reached.
    • The process will stop automatically and position the tool at the end point
      with Z ~50 μm above final Z depth when the set time has past.



    Expand
    titleClick here to expand Processing Stage 4 - Final polish

    Processing Stage 4 - Final polish

    InstructionsGraphics
    Step 1 - Raise the head to allow access to the sample and tool.

    Step 2 - Thoroughly clean your sample and realign for intermediate polish

    • After the blue compound it is necessary to thoroughly clean your
      sample to remove large particles that would cause scratches that can't be
      removed during the intermediate polish step.
    • If the sample was removed for cleaning, you must realign and clamp your
      sample, see these instructions.

    Step 3 - Mount the selected diameter Xbove tool into the spindle.

    • Make sure to mount the tool reserved for the yellow diamond paste, it will
      be marked with yellow paint on the mounting flat.
    • Make sure to align the flat on the tool shank with the set screw
      when inserting the tool into the spindle.

    Step 4 - Apply one to two drops of Extender Fluid to the thinned area
    and spread yellow diamond paste.

    Step 5 - Gently lower the head into operating position.

      • Be careful, if you release the locking mechanism and allow the
        polishing head to drop onto the force sensor, damage to the tool
        will occur.

    Step 6 - Define the top surface of the sample as the origin, or zero,
    point for the Z axis.

    • Since you have changed tools, we must re establish the Z origin.
    • Make sure force sensor is on.
    • Make sure is displayed before continuing. If
      is not displayed, click to switch joystick control to the
      Z axis.
    • Using the joystick bring the tool down to the sample surface until
      100 gm is reached. Z motion should stop automatically at 100 gm.
      Displayed on real time camera view.
    • Click to navigate to the Keypad Screen.
    • Click and then . The display on real time camera
      view will update to show Z position as 0.
    • Click , enter 50 into field, then click . The head will
      lift 50 μm off the sample and force should read ~0.

    Step 7 - Move to the center of the selected area

    • Navigate to the Set Scan Start/Endpoints screen
    • Press . The head will move to the defined XY origin.

    Step 8 - Change the timer for the polishing process if applicable. If no
    change needed, skip to next step.
    (600 seconds in this example)

    • Navigate to the Keypad screen.
    • Click , enter processing time into the field, then click .
      This will not produce an immediate and noticeable result, however it does
      set the Run timer shown on the Run screen.
    • Follow this step's instructions to reset the timer. Setting the timer,
      even to the same value, resets the timer. However, the displayed elapse
      time will not change to 0 until the button is pressed on
      the Run screen.

    Step 9 - Start the spindle (2000 RPM recommended)

    • From the Keypad screen, click , enter desired rotational speed into
      the field, then click . The spindle will begin rotating at the
      set speed.

    Step 10 - Begin the Final polishing process

    • Navigate to the Run screen
    • You should see your entered timer value in the Timer section
      under Set(Sec).
    • Make sure the Timer box is checked, indicating the tool should stop when
      the Elapsed time equals Set time.
    • Click and the tool will start rastering across the selected
      area.
    • Click to navigate to the Keypad Screen.
    • Click , enter an unreachable depth into field as a negative value,
      then click . This ensures you are controlled by the timer.
    • While continuing to raster, the head will slowly move down onto the part
      until the set polishing force is reached.
    • The process will stop automatically and position the tool at the end point
      with Z ~50 μm above final Z depth when the set time has past.





    Questions & Troubleshooting



    Process Library


    References

    Manufacturer Website

    ASAP-1 IPS Data Sheet

    ASAP-1 Product Brochure

    ASAP-1IPS Rev1.74.pdf

    ASAP-1IPS Rev1.76.pdf

    SAP presentation.pptx

    processing time calc.xlsx

    Widget Connector
    urlhttps://www.youtube.com/watch?v=O7rpZgH0y9Q

    Widget Connector
    urlhttps://www.youtube.com/watch?v=F3cyXxYKPjM

    Widget Connector
    urlhttps://www.youtube.com/watch?v=wvxIs3-qAgs

    Widget Connector
    urlhttps://www.youtube.com/watch?v=mVqJLc3-_e8

    Widget Connector
    urlhttps://www.youtube.com/watch?v=LMbsIjKutNQ