The JEOL system is calibrated by staff every 7 - 21 days, with separate calibration necessary for each condition file (current). These are typically stable for 1-3 weeks. As part of this process, column shift and tilt is adjusted (similar to an SEM), current is measured (and adjusted via the column zoom lenses, if necessary), wobble is checked and minimized (via the objective aperture, again similar to an SEM), focus and astigmatism is adjusted (via the objective lens strength and stigmator correction values), and the DAILYCAL file is run to ensure everything passes. Data is shown for the current FEG and previous FEG. These are replaced approximately every 2 years. Some particularly relevant measurements are shown below:
High currents (>30 nA) have their focus and astigmatism set manually by staff rather than through the built-in calibration routines on the system. Similarly, the built-in measurement programs do not give reliable values for beam size or astigmatism. As a result, only beam current (which is measured similarly and reliably compared to low currents) is posted here. |
...