Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.


Page Properties
idFunctionality

Insert excerpt
Problem Reporting Guide
Problem Reporting Guide
nopaneltrue

StatusDOWN
Issue Date and Description

 Melted o-rings - investigating Tube cracked, sourcing replacement.

Estimated Fix Date and Comment

 Ordering replacement tube. 

Responding StaffR. Hosler


Insert excerpt
e-Log Link - Protemp Horizontal Furnaces
e-Log Link - Protemp Horizontal Furnaces
nopaneltrue

Protemp Horizontal Furnace - Tube 02 - Silicon Nitride Deposition

iLab Name: C - Tube 02 Nitride Dep
iLab Kiosk: BRK Furnace Core
FIC:
 Shared
Owner: Rich Hosler
Location:
Cleanroom - R Bay
Maximum Wafer Size: 
6"/150 mm (for all furnaces)


Table of Content Zone
locationtop
styledisc

Table of Contents
outlinetrue
indent25px
stylenone


Overview


General Description

Horizontal furnaces for a variety of processes.

Specifications

Process Capabilities:

  • Tube 02 - LPCVD Nitride Deposition

Process Temperatures:

  • Tube 02 - 800°C maximum

Technology Overview 


Standard Operating Procedure

Widget Connector
urlhttp://youtube.com/watch?v=6RNKkDN_fUY
 - CVD


Questions & Troubleshooting

How long can my recipe go for with this tube? I want to make a thick deposition.


The current maximum length at 800°C is approximately 76 minutes. If this is exceeded you risk melting the silicone o-ring holding the Ammonia and Dicholorosliane in. If you need to do depositions longer than 76 minutes you can re-run the cycle after it fully completes the first time, which gives the o-ring time to cool down.