Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.


Page Properties
idFunctionality

Insert excerpt
Problem Reporting Guide
Problem Reporting Guide
nopaneltrue

StatusDOWNUP
Issue Date and Description

 Tube cracked, sourcing replacement.


Estimated Fix Date and Comment

 Ordering replacement tube. 


Responding StaffR. Hosler


Insert excerpt
e-Log Link - Protemp Horizontal Furnaces
e-Log Link - Protemp Horizontal Furnaces
nopaneltrue

Protemp Horizontal Furnace - Tube 02 - Silicon Nitride Deposition

iLab Name: C - Tube 02 Nitride Dep
iLab Kiosk: BRK Furnace Core
FIC:
 Shared
Owner: Rich Hosler
Location:
Cleanroom - R Bay
Maximum Wafer Size: 
6"/150 mm (for all furnaces)


Table of Content Zone
locationtop
styledisc

Table of Contents
outlinetrue
indent25px
stylenone


Overview


General Description

Horizontal furnaces for a variety of processes.

Specifications

Process Capabilities:

  • Tube 02 - LPCVD Nitride Deposition

Process Temperatures:

  • Tube 02 - 800°C maximum

Technology Overview 


Standard Operating Procedure

Widget Connector
urlhttp://youtube.com/watch?v=6RNKkDN_fUY
 - CVD


Questions & Troubleshooting

How long can my recipe go for with this tube? I want to make a thick deposition.


The current maximum length at 800°C is approximately 76 minutes. If this is exceeded you risk melting the silicone o-ring holding the Ammonia and Dicholorosliane in. If you need to do depositions longer than 76 minutes you can re-run the cycle after it fully completes the first time, which gives the o-ring time to cool down.