iLab Name: C - Tube 04 Field Ox
iLab Kiosk: BRK Furnace Core
FIC: Shared
Owner: Rich Hosler
Location: Cleanroom - R Bay
Maximum Wafer Size: 6"/150 mm (for all furnaces)
This tube has a damaged from heating zone. Tests were run to determine if this is causing lower rates.
Dry oxide rate was determined to be 7% lower than target on the default 200nm recipe.
Wet oxide rate was determined to be 36% lower than target on the default 1000nm recipe.
The tube is still functional and usable, just bear the above in mind with recipe lengths.