Page Properties | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| ||||||||||||||||
|
Insert excerpt | ||||||
---|---|---|---|---|---|---|
|
Protemp Horizontal Furnace - Tube 04 - Field Oxide Growth
iLab Name: C - Tube 04 Field Ox
iLab Kiosk: BRK Furnace Core
FIC: Shared
Owner: Rich Hosler
Location: Cleanroom - R Bay
Maximum Wafer Size: 6"/150 mm (for all furnaces)
This tube has a damaged from heating zone. Tests were run performed to determine if this is causing lower rates. (oct_23_ox_check.xlsx)
Dry oxide rate was determined to be 7% lower than target on the default 200nm recipe.
Wet oxide rate was determined to be 36% lower than target on the default 1000nm recipe.
The tube is still functional and usable, just bear the above in mind with recipe lengths.
Table of Content Zone | ||||||||
---|---|---|---|---|---|---|---|---|
| ||||||||
|
Overview
General Description
Horizontal furnaces for a variety of processes.
Specifications
Process Capabilities:
- Tube 04 - Wet / Dry Oxidation (For general use)
Process Temperatures:
- Tube 04 - 1100°C maximum
Technology Overview
Sample Requirements and PreparationStandard Operating Procedure
Widget Connector | ||
---|---|---|
|