Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.


Page Properties
idFunctionality

Insert excerpt
Problem Reporting Guide
Problem Reporting Guide
nopaneltrue

StatusShutdown - Longterm
Issue Date and Description


Estimated Fix Date and Comment
Controller board issue
Responding StaffTim Miller




iLab Name: K and S 4526
iLab Kiosk: BRK Packaging and Assembly Core
FIC:
David Janes
Owner: Tim Miller
Location:
BRK 2261
Maximum Wafer Size: 


Table of Content Zone
locationtop
styledisc

Table of Contents
outlinetrue
indent25px
stylenone


Overview

General Description

These systems use gold wire to bond. The pad surfaces to be bonded must be compatible with gold or the bonds will not stick. Gold, copper, anuminum, or silver work well. Titanium, ITO, and diamond can not be wire bonded

Specifications

The West Bond bonder has 25 micron gold wire and is operated manually. The K&S bonder has a wire size of 125 microns by 25 microns and may be operated manually or by programming. The pads must be larger than the size of a wire bond foot. Wire bond feet are typically 50 microns in width and 75 microns long. The recommended pads are 100 x100 microns minimum size on both chip and package for the West Bond. 100 x 150 microns for the K&S Ribbon Bonder.

Technology Overview - Remove if multiple tools use the same technology/process

 

 

Sample Requirements and Preparation

 

Standard Operating Procedure


Questions & Troubleshooting


 

Process Library


References