Overview


General Description

Horizontal furnaces for a variety of processes.

Specifications

Process Capabilities:

  • Tube 02 - LPCVD Nitride Deposition

Process Temperatures:

  • Tube 02 - 800°C maximum

Technology Overview 


Standard Operating Procedure

Widget Connector
urlhttp://youtube.com/watch?v=6RNKkDN_fUY
 - CVD


Questions & Troubleshooting

How long can my recipe go for with this tube? I want to make a thick deposition.


The current maximum length at 800°C is approximately 76 minutes. If this is exceeded you risk melting the silicone o-ring holding the Ammonia and Dicholorosliane in. If you need to do depositions longer than 76 minutes you can re-run the cycle after it fully completes the first time, which gives the o-ring time to cool down.