Check moisture and O2 levels, regenerate / purge as necessary
Perform 1 min leak test. Should be < 0.1 vol%/hr.
Perform moisture analyzer maintenance
Gloves and glove O-rings should be changed once every (3) months or as needed based on the condition.
Gloves, Left/Right, 1 pair (Universal can be used in any Glovebox); 2 pair required
Glove O-Rings (4 per pair of gloves); 2 sets required
The inlet/outlet filters should be replaced every (6) months.
Inlet/Outlet HEPA Filter
The large antechamber door O-rings should be replaced every year.
Large Antechamber Door O-Ring (2 per door)
The small antechamber door O-rings should be replaced every year.
Small Antechamber Door O-Ring (2 per door)
Check for Alarm Signal on Pump
Check Oil level
C-0402-000-0111
BARRIERRTA J100ES Lubricant oil