Check to ensure the instrument is in the proper state before you start.
- The bell jar should be fully lowered onto the base.
- The ion gauge controller should be powered on, and indicate a pressure range of X.X - 08 or below.
![](https://purdue.atlassian.net/wiki/download/thumbnails/6239234/image2019-5-13_16-3-58.png?version=1&modificationDate=1557777838000&cacheVersion=1&api=v2&width=200&height=53)
- The TELEMARK crucible indexer should be powered on and indicate which source is positioned for deposition.
![](https://purdue.atlassian.net/wiki/download/thumbnails/6239234/image2019-5-13_16-4-43.png?version=1&modificationDate=1557777884000&cacheVersion=1&api=v2&width=200&height=40)
- The INFICON depositoin controller should be powered on and indicate READY
![](https://purdue.atlassian.net/wiki/download/thumbnails/6239234/image2019-5-13_16-5-23.png?version=1&modificationDate=1557777923000&cacheVersion=1&api=v2&width=200&height=88)
- The TELEMARK CHEETAH programmable sweep controller should be powered on...?
![](https://purdue.atlassian.net/wiki/download/thumbnails/6239234/image2019-5-13_16-6-2.png?version=1&modificationDate=1557777963000&cacheVersion=1&api=v2&width=200&height=42)
- The M.A.P.S. HTR Power Supply should be powered off.
![](https://purdue.atlassian.net/wiki/download/thumbnails/6239234/image2019-5-13_16-6-36.png?version=1&modificationDate=1557777996000&cacheVersion=1&api=v2&width=150&height=92)
- The TELEMARK TT-6 power supply should be off, with the main breaker in the down position